OMCL Network of the Council of
Europe
QUALITY MANAGEMENT
DOCUMENT
质量管理文件
PA/PH/OMCL (10) 86 2R
QUALIFICATION OF EQUIPMENT
仪器确认
ANNEX 7:
QUALIFICATION OF MASS SPECTROMETERS
附件<
/p>
7
:质谱仪的确认
Full document title and
reference
文件全名和索引号
Document
type
文件类型
Legislative basis
立法基础
Date of
first adoption
首次发行日期
Date of
original entry into force
首次执行日期
Date of
entry into force of revised
document
修订后执行日期
Previous
titles/other references
原文件
名
/
其它索引号
Custodian Organisation
保管机构
Concerned
Network
相关网络
Qualification of Equipment
Annex 7: Qualification of mass
spectrometers
PA/PH/OMCL(10)86 2R
Guideline
-
25
th
May 2011
1
st
July 2011
-
-
The present
document was elaborated by the OMCL
Network/EDQM of the Council of Europe
GEON
ANNEX 7 OF THE OMCL
NETWORK GUIDELINE
“QUALIFICATION OF EQUIPMENT”
OMCL
网络指南“仪器确认”附件
7
7.1: QUALIFICATION OF GAS
CHROMATOGRAPHY-MASS SPECTROMETERS
WITH
ELECTRON IMPACT IONIZATION (GC-EI-MS)
装
备电子碰撞电离的气相色
谱
-
质谱仪的
确认
Introduction
介绍
The present
document is the first part of the 7
th
Annex of the co
re document
“Qualification of
Equipment”, and it
should be used in combination with it when
planning, performing and
documenting
the qualification process of GC-EI-MS.
本文件是核心文件“仪器确认”的第
7
个附件,在计实施和记录
GC-EI-MS
的确认过程中,
p>
应与核心文件结合使用。
The
core document contains the Introduction and
general forms for Level I and II of
qualification, which are common to all
type of instruments.
核心文件包括
了介绍和适用于所有类型仪器的第一级和第二级确认用通用表格。
The present annex contains instrument-
related recommendations on parameters to be
checked
at Level III and IV of
qualification and the corresponding typical
acceptance limits, as well as
practical
examples on the methodology that can be used to
carry out these checks.
本附件
包括与仪器相关的在第三级和第四级要检查的参数及其相关的典型可接受限度的
推荐,以
及在实施这些检查时可以采用的方法学实例。
TABLE
III
表
3
Level III.
Periodic and motivated instrument checks
第三级
周期性主动仪检查
Minimum
requirements for GC-EI-MS
(examples)
GC-EI-MS
最低要求(举例)
Parameter to be checked
要检查的参数
Mass
accuracy
质量准确性
PFTBA (FC-43) *
Or internal
calibration gas
或内部校正气体
Linearity
**
线性
System/instrument precision **
系统
/
仪器
精密度
*)
PFTBA (FC-43): Perfluoro-tributyl-amine
(CAS NO.: 311-89-7)
Typical tolerance
limits
典型允许限度
m/z
= 69±
0.5
m/z =
219±
0.5
m/z =
502±
0.5
or defined masses of
internal calibration gas ±
0.5
或内校气体指定的质量
±
0.5
r2 ≥0.995
≤1.0%
如果要进行定量
PFTBA(FC-43):
全氟三丁胺(
< br>CAS
号
311-89-7
)
**)
In case of
quantification purposes, these parameters have to
be checked.
检测,这些参数必须要检查。
TABLE IV
表四
Level IV. In-use instrument checks
第四级
在用仪器的检查
Requirements for GC-EI-MS
GC-EI-MS
的要求
Parameter to be checked
要检查的参数
As defined
by the specific analysis method or
according to the Ph. Eur. Or the MAH
dossier
(see example in Level
IV
. Annex II)
如特定分析方法定义,或根据欧
洲药典或
MAH
档案(见第四级附件
2
的举例)
Typical tolerance limits
典型允许限度
Identified
by library or with reference standard
由图谱库进行鉴别或与对照品进行对照
ANNEX I
附件
1
Level III. Periodic and motivated
instrument checks
第三级
周期性主动仪器检查
This
Annex contains
practical examples
of tests and their associated tolerance
limits for
several parameters related
to the performance of GC-EI-MS.
本附件包括了
GC-EI-
MS
性能相关的几个参数的测试实例及其关联的允许限度。
These
examples
can be considered by the OMCLs as
possible approaches to perform the Level
III of the equipment qualification
process: “Periodic and motivated instrument
checks”.
这些实例可以作为
OMCL
在实施第三级仪器确认过程“周期性主动仪器检查”中可以采
用的方法。
GENERAL
CONSIDERATIONS
一般考虑
- GC-MS is mainly used for the
identification of unknown substances or
quantification of low
concentrated
substances where high specificity is needed.
GC-MS
主要用于需要较高专属性
的未知物质的鉴别,或低浓度物质的定量
MASS
ACCURACY
质量准确度
Materials:
材料
PFTBA (FC-43) or internal calibration
gas
PFTBA(FC-43)
或内部校正气体
Method:
方法
Internal
instrument check or spectrum of PFTBA (FC-43) in
full scan mode
仪器内部检查或
PFTBA
(
FC-43
)全扫描模式光谱扫描
Limits:
限度
m/z = 69
±
0.5
m/z = 219
±
0.5
m/z = 502
±
0.5
Note
: the tolerance limits
for mass accuracy are only valid for quadrupole
and ion trap mass
spectrometers.
Instruments like Time-of-Flight (TOF) or hybrid
instruments like
quadrupole-Time-of-
Flight (QTOF) have a much better mass accuracy and
adequate
requirements should be applied
by the laboratory.
注:质量准确度的
允许限度只对四级杆和离子阱质谱仪有效。其它仪器,如飞行时间质
谱
< br>(
TOF
)
或
< br>HYBRID
仪器,
如四级杆
-
飞行时间质谱
(
QTOF
)
,
具有更好的质量准确性,
这时化验室应采用足够好的要求。
LINEARITY
线性
Materials:
材料
Stock solutions: 1-
Octanol
in dichloromethane 0.2, 0.4, 0.6, 0.8, 1.0 μL/mL
贮备液:
0.2, 0.4,
0.6, 0.8, 1.0
μL/mL
正辛醇的二氯甲烷溶液
Method:
方法
2 injections of each level (injection
volume: 1.0 μL)
每个水平
2
针(进样量
1.0uL
)
Limits:
限度
r? ≥ 0.995
SYSTEM/INSTRUMENT PRECISION
系统
/
仪器精密度
Materials:
材料
Stock solution: 1-
Octanol in
dichloromethane 1.0 μL/mL
贮备液:
1.0
μL/mL
正辛醇的二氯甲烷溶液
Method:
方法
6 injections
(injection
volume: 1.0 μL)
6
针(进样量:
1.0uL
)
Limits (minimum requirement):
限度(最低要求)
RSD ≤
10.0% (without internal standard
没有内控标准
)
-
-
-
-
-
-
-
-
-