3 General requirements for the performance


evaluation


3.1 Responsibilities and resources


The manufacturer takes the responsibility for the initiation and/or the


conduct of a performance evaluation study. He shall define the


responsibility and the interrelation of all personnel who manage


and conduct the performance evaluation of IVD MDs, particularly for


personnel who need the


organisational freedom and authority to


a) assess the validity of test results and data already available;


b) specify performance claims which shall be further examined or


confirmed;


c) specify and document the evaluation plan and the test procedures;


d) prepare the evaluation report.


The manufacturer shall appoint a co-ordinator with overall


responsibility of the performance evaluation


study. The co-ordinator shall himself assure that adequate resources


are available. The investigator


shall ensure that the evaluation plan is followed at his location and


that the study is appropriately


reviewed from an ethical point of view.


3.2 Documentation


The documentation of the performance evaluation study shall contain


the files relating to clauses 3 to


7 of this standard and shall be part of the technical documentation of


the IVD MD.


3.3 Final assessment and review


The co-ordinator shall assess and document which performance


claims are met, state whether claims


are not met and give recommendations for corrective actions, where


necessary.


The responsible management of the manufacturer shall make sure


that the results of the performance


evaluation study and the recommendations for corrective actions are


carefully considered and properly


documented before issuing a declaration of conformity.


4 Organisation of a performance evaluation study


4.1 Preconditions


Before starting a performance evaluation study it shall be ensured by


the co-ordinator that


a) the performance claims of the IVD MD which are the subject of the


study are specified;


b) the IVD MD has been manufactured under controlled production


processes and conditions;


c) the IVD MD to be evaluated meets the quality control release


specifications;


d) a sufficient number of samples of the IVD MD can be provided


during the entire period of the


performance evaluation study;


e) all legal requirements for performance evaluation studies are met;


f) the investigator(s) is (are) adequately skilled and trained to conduct


the study and the necessary


resources are available.


4.2 Evaluation plan


The evaluation plan shall state the purpose on scientific, technical or


medical grounds, the scope of


the evaluation, the structure and organization of the study and the


number of devices concerned.


Defining the objective of the study, the co-ordinator shall have


assessed which performance claims


are already verified by data or scientific literature.



The evaluation plan shall be designed to minimise the requirements


for invasive sampling. In the case


of IVD MDs for self-testing it shall be ensured that the evaluation plan


is appropriate and acceptable to


users and the information provided shall be clear and easily


understood.


The evaluation plan shall specify


a) that the investigator(s) is (are) adequately skilled and trained to


use the IVD MD;


b) the list of laboratories or other institutions taking part in the


performance evaluation study; for selftesting,


the location and number of lay persons involved;


c) the time- table;


d) the necessary minimum number of probands from whom


specimens are collected by invasive


procedures in order to adequately assess the performance of the IVD


MD;


e) instructions for use including a description of the conditions of use;


f) the performance claims (e.g. analytical sensitivity, diagnostic


sensitivity, analytical specificity,


diagnostic specificity, accuracy, repeatability, reproducibility) to be


validated;


g) the format of performance study records.


4.3 Sites and resources


In general, the performance study procedure(s) shall be carried out


under conditions reflecting the


relevant intended conditions of use.


The co-ordinator shall take the responsibility for the proper conduct of


the performance evaluation


study at all sites. All investigators shall be named.


The co-ordinator shall ensure adequate competence and skill at all


sites involved and that the


necessary resources are available.


Where lay persons are involved in a performance evaluation study of


an IVD MD for self-testing, the


location of the study and the number of persons shall be given. The


co-ordinator shall specify the


criteria for the selection of a representative panel.


Especially for studies involving lay persons it shall be ensured that


these persons do not receive


additional information on the use of the IVD MD apart from that which


is provided with the IVD MD


when it is placed on the market because the comprehension of the


manufacturer's instructions for use


is one of the important aspects of the study. lt shall also be ensured


that the untrained person(s) do


not receive any additional information or help, e. g. from a tutor, other


than the training specified and


provided by the manufacturer in the instructions for use.


4.4 Basic design information


The co-ordinator shall provide the investigator(s) with sufficient


information in order to understand the


function and application of the IVD MD and, where necessary, the


investigator shall make himself


familiar with the IVD MD and its application. The information provided


shall include a statement that


the device in question conforms with the requirements of the Directive


98/79/EC apart from those to be


evaluated.



4.5 Experimental design


The experimental procedures to validate each performance claim


subject to the performance


evaluation study shall be documented in the evaluation plan.


Special consideration in performance evaluation studies of


reagents/kits shall be given, where


applicable, to the following:



specification of type (e.g. serum, plasma, urine) and properties (e.g.


concentration range, age


and sex of the proband population) of specimens appropriate to the


intended use;



probands to be enrolled;



suitability, stability and volume of specimens and specimen


exclusion criteria;




blind procedures, where necessary;



reagent stability;



inclusion of common interfering factors, caused by specimen


condition or the


pathological/physiological status of the specimen donor or treatment;



conditions for use which can be reasonably anticipated; special


attention shall be paid to the


conditions of use by lay persons;



selection of an appropriate reference measurement procedure and


reference material of


higher order, where available;



determination of the status of specimens (for qualitative tests with a


nominal or ordinal scale);



calibration procedures, including traceability, where appropriate;



appropriate means of control;



limitations of the test;



criteria for re- examination and data exclusion;



availability of additional information concerning the specimen or


donor if follow-up of


unexpected results is required;



appropriate measures to reduce risk of infection to the user.


Where the study is intended to validate the performance claims of an


instrument special consideration


shall be given additionally to the following:



maintenance and cleaning;



carry-over effects;



software validation.


NOTE For the investigation of the technical aspects of instruments, other


standards can be relevant.


4.6 Performance study records


The performance study records shall



refer to the experimental procedures in the evaluation plan;



be unequivocally identifiable;



contain or refer to all results and related relevant data;



be part of the technical documentation of the IVD MD.


The protection of all confidential data shall be ensured.



4.7 Observations and unexpected outcomes


Special attention shall be paid to observations and unexpected


outcomes, e. g. drop outs, outliers,


instability of sample or reagent signal etc., non-reproducibility, non-


correlation of results to the


reference or to the diagnostic pattern, defects or breakdowns,


software errors, or error signals.


Any deviation from the defined procedures shall be recorded. In the


case of IVD MDs for self-testing,


the investigator or tutor shall duly note any difficulty or question a


user may have and any deviation


from the mode of application of the IVD MD as described by the


manufacturer.


Any such observation shall be properly recorded. The co- ordinator


shall, together with the investigator,


trace the cause whenever possible. The result shall be recorded and


shall be part of the evaluation


report.


Where the validity of the examinations already performed may be


questionable because of an


identified source of error the tests shall be repeated after exclusion of


that cause.


Where a misuse or misinterpretation of the instructions for use has


been the cause and where an


unexpected risk inherent to the product design or the mode of


application has been identified this shall


be clearly stated.


The proposals of the investigator(s) and the co-ordinator for any


improvement of the IVD MD and/or its


application shall be recorded.


4.8 Evaluation report


The co-ordinator shall establish an evaluation report. It shall contain a


description of the study, an


analysis of the results together with a conclusion on the performance


claims investigated.


The report shall also discuss any unexpected outcomes which have


occurred. It shall identify the


cause whenever possible and give recommendations for corrective


actions to be taken, where


necessary.


If several studies have been conducted for one IVD MD, a single


summarizing report may be


established.


5 Modifications during the performance


evaluation study


Where the manufacturing process has been changed it shall be


checked whether the performance


claims of the IVD MD still conform to those which had been set


initially. Otherwise the validity of the


examinations already performed shall be questioned and the


evaluation plan shall be revised


accordingly.


Where design changes are introduced, the evaluation plan shall be


revised.


6 Re-evaluation


In case of changes to the design or manufacturing process of the IVD


MD, the performance evaluation


study shall be repeated as far as necessary, to ensure that the


intended use and the performance


claims of the IVD MD placed on the market are adequately evaluated.

-


-


-


-


-


-


-


-



本文更新与2021-02-18 13:12,由作者提供,不代表本网站立场,转载请注明出处:https://www.bjmy2z.cn/gaokao/665281.html

体外诊断试剂IVD欧盟EN13612性能评估部分中英文的相关文章

  • 余华爱情经典语录,余华爱情句子

    余华的经典语录——余华《第七天》40、我不怕死,一点都不怕,只怕再也不能看见你——余华《第七天》4可是我再也没遇到一个像福贵这样令我难忘的人了,对自己的经历如此清楚,

    语文
  • 心情低落的图片压抑,心情低落的图片发朋友圈

    心情压抑的图片(心太累没人理解的说说带图片)1、有时候很想找个人倾诉一下,却又不知从何说起,最终是什么也不说,只想快点睡过去,告诉自己,明天就好了。有时候,突然会觉得

    语文
  • 经典古训100句图片大全,古训名言警句

    古代经典励志名言100句译:好的药物味苦但对治病有利;忠言劝诫的话听起来不顺耳却对人的行为有利。3良言一句三冬暖,恶语伤人六月寒。喷泉的高度不会超过它的源头;一个人的事

    语文
  • 关于青春奋斗的名人名言鲁迅,关于青年奋斗的名言鲁迅

    鲁迅名言名句大全励志1、世上本没有路,走的人多了自然便成了路。下面是我整理的鲁迅先生的名言名句大全,希望对你有所帮助!当生存时,还是将遭践踏,将遭删刈,直至于死亡而

    语文
  • 三国群英单机版手游礼包码,三国群英手机单机版攻略

    三国群英传7五神兽洞有什么用那是多一个武将技能。青龙飞升召唤出东方的守护兽,神兽之一的青龙。玄武怒流召唤出北方的守护兽,神兽之一的玄武。白虎傲啸召唤出西方的守护兽,

    语文
  • 不收费的情感挽回专家电话,情感挽回免费咨询

    免费的情感挽回机构(揭秘情感挽回机构骗局)1、牛牛(化名)向上海市公安局金山分局报案,称自己为了挽回与女友的感情,被一家名为“实花教育咨询”的情感咨询机构诈骗4万余元。

    语文