-
621
CHROMATOGRAPHY
INTRODUCTION
This chapter defines the
terms and procedures used in chromatography and
provides
general information. Specific
requirements for chromatographic procedures for
drug
substances and dosage forms,
including adsorbent and developing solvents, are
given in the individual monographs.
Chromatography is defined as a
procedure by which solutes are separated by a
dynamic differential migration process
in a system consisting of two or more phases,
one of which moves continuously in a
given direction and in which the individual
substances exhibit different mobilities
by reason of differences in adsorption, partition,
solubility, vapor pressure, molecular
size, or ionic charge density. The individual
substances thus separated can be
identified or determined by analytical procedures.
The general chromatographic technique
requires that a solute undergo distribution
between two phases, one of them fixed
(stationary phase), the other moving (mobile
phase). It is the mobile phase that
transfers the solute through the medium until it
eventually emerges separated from other
solutes that are eluted earlier or later.
Generally, the solute is transported
through the separation medium by means of a
flowing stream of a liquid or a gaseous
solvent known as the ―eluant.‖ The stationary
phase may act through adsorption, as in
the case of adsorbents such as activated
alumina and silica gel, or it may act
by dissolving the solute, thus partitioning the
latter
between the stationary and
mobile phases. In the latter process, a liquid
coated onto
an inert support, or
chemically bonded onto silica gel, or directly
onto the wall of a
fused silica
capillary, serves as the stationary phase.
Partitioning is the predominant
mechanism of separation in
gas
–
liquid chromatography,
paper chromatography, in
forms of
column chromatography, and in thin-layer
chromatography designated as
liquid-
liquid chromatography. In practice, separations
frequently result from a
combination of
adsorption and partitioning effects. Other
separation principles include
ion
exchange, ion-pair formation, size exclusion,
hydrophobic interaction, and chiral
recognition.
The types of
chromatography useful in qualitative and
quantitative analysis that are
employed
in the
USP
procedures are
column, gas, paper, thin-layer, (including
high-performance thin-layer
chromatography), and pressurized liquid
chromatography
(commonly called high-
pressure or high-performance liquid
chromatography). Paper
and thin-layer
chromatography are ordinarily more useful for
purposes of identification,
because of
their convenience and simplicity. Column
chromatography offers a wider
choice of
stationary phases and is useful for the separation
of individual compounds,
in quantity,
from mixtures. Modern high-performance thin-layer
chromatography, gas
chromatography, and
pressurized liquid chromatography require more
elaborate
apparatus but usually provide
high resolution and identify and quantitate very
small
amounts of material.
Use of Reference Substances in Identity
Tests
—
In paper and thin-
layer
chromatography, the ratio of the
distance (this distance being measured to the
point of
maximum intensity of the spot
or zone) traveled on the medium by a given
compound
to the distance traveled by
the front of the mobile phase, from the point of
application
of the test substance, is
designated as the
R
F
value of the
compound. The ratio
between the
distances traveled by a given compound and a
reference substance is
the
R
R
value.
R
F
values vary
with the experimental conditions, and thus
identification is
best accomplished
where an authentic specimen of the compound in
question is used
as a reference
substance on the same chromatogram.
For this purpose, chromatograms are
prepared by applying on the thin-layer
adsorbent or on the paper in a straight
line, parallel to the edge of the
chromatographic plate or paper,
solutions of the substance to be identified, the
authentic specimen, and a mixture of
nearly equal amounts of the substance to be
identified and the authentic specimen.
Each sample application contains
approximately the same quantity by
weight of material to be chromatographed. If the
substance to be identified and the
authentic specimen are identical, all
chromatograms agree in color and
R
F
value and the
mixed chromatogram yields a
single
spot; i.e.,
R
R
is
1.0.
Location and Identification of
Components
—
The spots
produced by paper or
thin-layer
chromatography may be located by: (1) direct
inspection if the compounds
are visible
under white or either short-wavelength (254 nm) or
long-wavelength (360
nm) UV light, (2)
inspection in white or UV light after treatment
with reagents that will
make the spots
visible (reagents are most conveniently applied
with an atomizer), (3)
use of a Geiger-
Mü
ller counter or autoradiographic
techniques in the case of the
presence
of radioactive substances, or (4) evidence
resulting from stimulation or
inhibition of bacterial growth by the
placing of removed portions of the adsorbent and
substance on inoculated media.
In open-column
chromatography, in pressurized liquid
chromatography performed
under
conditions of constant flow rate, and in gas
chromatography, the retention time,
t,
defined as the time elapsed between
sample injection and appearance of the peak
concentration of the eluted sample
zone, may be used as a parameter of
identification.
Solutions of the
substance to be identified or derivatives thereof,
of the reference
compound, and of a
mixture of equal amounts of these two are
chromatographed
successively on the
same column under the same chromatographic
conditions. Only
one peak should be
observed for the mixture. The ratio of the
retention times of the
test substance,
the reference compound, and a mixture of these, to
the retention time
of an internal
standard is called the relative retention time
R
R
and is also
used
frequently as a parameter of
identification.
The deviations of
R
R
,
R
F
, or
t
values measured for the
test substance from the values
obtained
for the reference compound and mixture should not
exceed the reliability
estimates
determined statistically from replicate assays of
the reference compound.
Chromatographic
identification by these methods under given
conditions strongly
indicates identity
but does not constitute definitive identification.
Coincidence of
identity parameters
under three to six different sets of
chromatographic conditions
(
temperatures, column
packings, adsorbents, eluants, developing
solvents, various
chemical derivatives,
etc
.) increases the probability that
the test and reference
substances are
identical. However, many isomeric compounds cannot
be separated.
Specific and pertinent
chemical, spectroscopic, or physicochemical
identification of
the eluted component
combined with chromatographic identity is the most
valid
criterion of identification. For
this purpose, the individual components separated
by
chromatography may be collected for
further identification.
PAPER CHROMATOGRAPHY
In paper chromatography the
adsorbent is a sheet of paper of suitable texture
and
thickness. Chromatographic
separation may proceed through the action of a
single
liquid phase in a process
analogous to adsorption chromatography in columns.
Since
the natural water content of the
paper, or selective imbibition of a hydrophilic
component of the liquid phase by the
paper fibers, may be regarded as a stationary
phase, a partitioning mechanism may
contribute significantly to the separation.
Alternatively, a two-phase system may
be used. The paper is impregnated with one of
the phases, which then remains
stationary (usually the more polar phase in the
case
of unmodified paper). The
chromatogram is developed by slow passage of the
other,
mobile phase over the sheet.
Development may be ascending, in which case the
solvent is carried up the paper by
capillary forces, or descending, in which case the
solvent flow is also assisted by
gravitational force.
Differences in the
value of
R
F
have
been reported where chromatograms developed in
the direction of the paper grain
(machine direction) are compared with others
developed at right angles to the grain;
therefore, the orientation of paper grain with
respect to solvent flow should be
maintained constant in a series of chromatograms.
(The machine direction is usually
designated by the manufacturer on packages of
chromatography paper.)
Descending Chromatography
In descending chromatography, the
mobile phase flows downward on the
chromatographic sheet.
Apparatus
—
The
essential equipment for descending chromatography
consists of the
following:
A
vapor-tight chamber
provided with inlets for addition of
solvent or for releasing
internal
pressure. The chamber is constructed preferably of
glass, stainless steel, or
porcelain
and is so designed as to permit observation of the
progress of the
chromatographic run
without opening of the chamber. Tall glass
cylinders are
convenient if they are
made vapor-tight with suitable covers and a
sealing compound.
A
rack of
corrosion-resistant material
about 5 cm
shorter than the inside height of the
chamber. The rack serves as a support
for solvent troughs and for antisiphon rods
which, in turn, hold up the
chromatographic sheets.
One or more
glass troughs
capable of
holding a volume of solvent greater than that
needed for one chromatographic run. The
troughs must also be longer than the width
of the chromatographic sheets.
Heavy glass antisiphon rods
to be supported by the rack and running
outside of,
parallel to, and slightly
above the edge of the glass trough.
Chromatographic sheets
of
special filter paper at least 2.5 cm wide and not
wider than
the length of the troughs
are cut to a length approximately equal to the
height of the
chamber. A fine pencil
line is drawn horizontally across the filter paper
at a distance
from one end such that,
when the sheet is suspended from the antisiphon
rods with
the upper end of the paper
resting in the trough and the lower portion
hanging free
into the chamber, the line
is located a few centimeters below the rods. Care
is
necessary to avoid contaminating the
filter paper by excessive handling or by contact
with dirty surfaces.
Procedure
—
The
substance or substances to be analyzed are
dissolved in a suitable
solvent.
Convenient volumes, delivered from suitable
micropipets, of the resulting
solution,
normally containing 1 to 20 ?
g of the
compound, are placed in 6- to 10-mm
spots not less than 3 cm apart along
the pencil line. If the total volume to be applied
would produce spots of a diameter
greater than 6 to 10 mm, it is applied in separate
portions to the same spot, each portion
being allowed to dry before the next is added.
The spotted chromatographic
sheet is suspended in the chamber by use of the
antisiphon rod, which holds the upper
end of the sheet in the solvent trough. The
bottom of the chamber is covered with
the prescribed solvent system. Saturation of
the chamber with solvent vapor is
facilitated by lining the inside walls with paper
that is
wetted with the prescribed
solvent system. It is important to ensure that the
portion of
the sheet hanging below the
rods is freely suspended in the chamber without
touching
the rack or the chamber walls
or the fluid in the chamber. The chamber is sealed
to
allow equilibration (saturation) of
the chamber and the paper with the solvent vapor.
Any excess pressure is released as
necessary. For large chambers, equilibration
overnight may be necessary.
A volume of the mobile phase in excess
of the volume required for complete
development of the chromatogram is
saturated with the immobile phase by shaking.
After equilibration of the chamber, the
prepared mobile solvent is introduced into the
trough through the inlet. The inlet is
closed and the mobile solvent phase is allowed to
travel the desired distance down the
paper. Precautions must be taken against
allowing the solvent to run down the
sheet when opening the chamber and removing
the chromatogram. The location of the
solvent front is quickly marked, and the sheets
are dried.
The chromatogram
is observed and measured directly or after
suitable development
to reveal the
location of the spots of the isolated drug or
drugs. The paper section(s)
predetermined to contain the isolated
drug(s) may be cut out and eluted by an
appropriate solvent, and the solutions
may be made up to a known volume and
quantitatively analyzed by appropriate
chemical or instrumental techniques. Similar
procedures should be conducted with
various amounts of similarly spotted reference
standard on the same paper in the
concentration range appropriate to prepare a valid
calibration curve.
Ascending
Chromatography
In ascending
chromatography, the lower edge of the sheet (or
strip) is dipped into the
mobile phase
to permit the mobile phase to rise on the
chromatographic sheet by
capillary
action.
Apparatus
—
The
essential equipment for ascending chromatography
is substantially
the same as that
described under
Descending
Chromatography.
Procedure
—
The
test materials are applied to the chromatographic
sheets as directed
under
Descending Chromatography,
and above the level to which the paper
is dipped
into the developing solvent.
The bottom of the developing chamber is covered
with the
developing solvent system. If
a two-phase system is used, both phases are added.
It
is also desirable to line the walls
of the chamber with paper and to saturate this
lining
with the solvent system. Empty
solvent troughs are placed on the bottom of the
chamber, and the chromatographic sheets
are suspended so that the end on which
the spots have been added hangs free
inside the empty trough.
The chamber is sealed, and
equilibration is allowed to proceed as described
under
Descending Chromatography.
Then the developing solvent (mobile
phase) is added
through the inlet to
the trough in excess of the solvent required for
complete
moistening of the
chromatographic sheet. The chamber is resealed.
When the solvent
front has reached the
desired height, the chamber is opened and the
sheet is
removed and dried.
Quantitative analyses of the spots may
be conducted as described under
Descending
Chromatography.
THIN-LAYER
CHROMATOGRAPHY
In thin-layer chromatography, the
adsorbent is a relatively thin, uniform layer of
dry,
finely powdered material applied
to a glass, plastic, or metal sheet or plate,
glass
plates
being most commonly employed. The
coated plate can be considered an ―open
chromatographic column‖ and the
separations achieved may be based upon
adsorption, partition, or a combination
of both effects, depending on the particular
type of stationary phase, its
preparation, and its use with different solvents.
Thin-layer
chromatography on ion-
exchange layers can be used for the fractionation
of polar
compounds. Presumptive
identification can be effected by observation of
spots or
zones of identical
R
F
value and
about equal magnitude obtained, respectively, with
an
unknown and a reference sample
chromatographed on the same plate. A visual
comparison of the size or intensity of
the spots or zones may serve for
semiquantitative estimation.
Quantitative measurements are possible by means of
densitometry (absorbance or
fluorescence measurements), or the spots may be
carefully removed from the plate,
followed by elution with a suitable solvent and
spectrophotometric measurement. For
two-dimensional thin-layer chromatography,
the chromatographed plate is turned at
a right angle and again chromatographed,
usually in another chamber equilibrated
with a different solvent system.
Apparatus
—
Acceptable apparatus and materials for thin-layer
chromatography
consist of the
following.
A
TLC
or HPTLC plate.
The chromatography is
generally carried out using
precoated
plates
or
sheets
(on glass, aluminum,
or polyester support) of suitable size. It may be
necessary to clean the plates prior to
separation. This can be done by migration of, or
immersion in, an appropriate solvent.
The plates may also be impregnated by
procedures such as development,
immersion, or spraying. At the time of use, the
plates may be activated, if necessary,
by heating in an oven at 120
for 20 minutes.
The
stationary phase
of TLC
plates has an average particle size of
10
–
15 ?
m, and
that of HPTLC plates an average
particle size of 5 ?
m. Commercial
plates with a
preadsorbant zone can be
used if they are specified in a monograph. Sample
applied
to the preabsorbant region
develops into sharp, narrow bands at the
preabsorbant-sorbent interface.
Alternatively, flat
glass
plates
of convenient size,
typically 20 cm ×
20 cm can
be coated as described under
Preparation of
Chromatographic Plates.
A suitable
manual, semiautomatic, or automatic
application device
can be used to
ensure proper positioning of the plate
and proper transfer of the sample, with respect
to volume and position, onto the plate.
Alternatively, a
template
can be used to guide
in manually
placing the test spots at definite intervals, to
mark distances as needed,
and to aid in
labeling the plates. For the proper application of
the solutions,
micropipets,
microsyringes, or calibrated disposable
capillaries
are recommended.
For ascending development, a
chromatographic chamber
made
of inert, transparent
material and
having the following specifications is used: a
flat bottom or twin trough, a
tightly
fitted lid, and a size suitable for the plates.
For horizontal development, the
chamber
is provided with a reservoir for the mobile phase,
and it also contains a
device for
directing the mobile phase to the stationary
phase.
Devices for transfer of
reagents
onto the plate by spraying,
immersion, or exposure to
vapor and
devices to facilitate any necessary heating for
visualization of the
separated spots or
zones.
A
UV light
source
suitable for observations under
short (254 nm) and long (365 nm)
wavelength UV light.
A
suitable
device for
documentation
of the visualized
chromatographic result.
Procedure
—
Apply
the prescribed volume of the test solution and the
standard
solution in sufficiently small
portions to obtain circular spots of 2 to 5 mm in
diameter
(1 to 2 mm on HPTLC plates) or
bands of 10 to 20 mm by 1 to 2 mm (5 to 10 mm by
0.5 to 1 mm on HPTLC plates) at an
appropriate distance from the lower
edge
—
during
chromatography the application position must be at
least 3 mm (HPTLC)
or 5 mm (TLC) above
the level of the developing
solvent
—
and from the sides
of the
plate. Apply the solutions on a
line parallel to the lower edge of the plate with
an
interval of at least 10 mm (5 mm on
HPTLC plates) between the centers of
spots
or 4
mm (2 mm on HPTLC
plates) between the edges of bands, and allow to
dry.
Ascending
Development
—
Line at least
one wall of the chromatographic chamber
with filter paper. Pour into the
chromatographic chamber a quantity of the mobile
phase sufficient for the size of the
chamber to give, after impregnation of the filter
paper, a level of depth appropriate to
the dimension of the plate used. For saturation
of the chromatographic chamber, close
the lid, and allow the system to equilibrate.
Unless otherwise indicated, the
chromatographic separation is performed in a
saturated chamber.
Place the plate in the chamber,
ensuring that the plate is as vertical as possible
and
that the spots or bands are above
the surface of the mobile phase, and close the
chamber. The stationary phase faces the
inside of the chamber. Remove the plate
when the mobile phase has moved over
the prescribed distance. Dry the plate, and
visualize the chromatograms as
prescribed. For two-dimensional chromatography,
dry the plates after the first
development, and carry out a second development in
a
direction perpendicular to that of
the first development.
Horizontal
Development
—
Introduce a
sufficient quantity of the developing solvent
into the reservoir of the chamber using
a syringe or pipet. Place the plate horizontally
in the chamber, connect the mobile
phase direction device according to the
manufacturer's instructions, and close
the chamber. If prescribed, develop the plate
starting simultaneously at both ends.
Remove the plate when the mobile phase has
moved over the distance prescribed in
the monograph. Dry the plate, and visualize the
chromatograms as prescribed.
For two-dimensional
chromatography, dry the plates after the first
development, and
carry out a second
development in a direction perpendicular to that
of the first
development.
Detection
—
Observe the dry plate first under short-wavelength
UV light (254 nm) and
then under long-
wavelength UV light (365 nm) or as stated in the
monograph. If
further directed, spray,
immerse, or expose the plate to vapors of the
specified
reagent, heat the plate when
required, observe, and compare the test
chromatogram
with the standard
chromatogram. Document the plate after each
observation.
Measure and record the
distance of each spot or zone from the point of
origin, and
indicate for each spot or
zone the wavelength under which it was observed.
Determine the
R
F
values for the
principal spots or zones (see
Glossary
of Symbols
).
Quantitative
Measurement
—
Using
appropriate instrumentation, substances
separated by TLC and responding to
ultraviolet-visible (UV-Vis) irradiation prior to
or
after derivatization can be
determined directly on the plate. While moving the
plate or
the measuring device, the
plate is examined by measuring the reflectance of
the
incident light. Similarly,
fluorescence may be measured using an appropriate
optical
system. Substances containing
radionuclides can be quantified in three ways: (1)
directly by moving the plate alongside
a suitable counter or vice versa; (2) by cutting
the plates into strips and measuring
the radioactivity on each individual strip using a
suitable counter; or (3) by scraping
off the stationary phase, dissolving it in a
suitable
scintillation cocktail, and
measuring the radioactivity using a liquid
scintillation counter
(see
Radioactivity
821
).
The apparatus for direct quantitative
measurement on the plate is a densitometer that
is composed of a mechanical device to
move the plate or the measuring device along
the
x
-axis and
the
y
-axis, a recorder, a
suitable integrator or a computer; and, for
substances responding to UV-Vis
irradiation, a photometer with a source of light,
an
optical device capable of generating
monochromatic light, and a photo cell of
adequate sensitivity, all of which are
used for the measurement of reflectance. In the
case where fluorescence is measured, a
suitable filter is also required to prevent the
light used for excitation from reaching
the photo cell while permitting the emitted light
or specific portions thereof to pass.
The linearity range of the counting device must be
verified.
For quantitative
tests, it is necessary to apply to the plate not
fewer than three
standard solutions of
the substance to be examined, the concentrations
of which span
the expected value in the
test solution (e.g., 80%, 100%, and 120%).
Derivatize with
the prescribed reagent,
if necessary, and record the reflectance or
fluorescence in the
chromatograms
obtained. Use the measured results for the
calculation of the amount
of substance
in the test solution.
Preparation of
Chromatographic Plates
—
Apparatus
—
Flat
glass
plates
of convenient size, typically 20
cm ×
20 cm.
An
aligning tray
or a flat
surface upon which to align and rest the plates
during the
application of the
adsorbent.
A
storage
rack
to hold the prepared plates during
drying and transportation. The rack
holding the plates should be kept in a
desiccator or be capable of being sealed in
order to protect the plates from the
environment after removal from the drying oven.
The
adsorbent
consists of finely divided adsorbent materials,
normally 5 to 40 ?
m in
diameter, suitable for chromatography.
It can be applied directly to the glass plate or
can be bonded to the plate by means of
plaster of Paris [calcium sulfate hemihydrate
(at a ratio of 5% to 15%)] or with
starch paste or other binders. The plaster of
Paris will
not yield as hard a surface
as will the starch, but it is not affected by
strongly oxidizing
spray reagents. The
adsorbent may contain fluorescing material to aid
in the
visualization of spots that
absorb UV light.
A
spreader,
which, when moved
over the glass plate, will apply a uniform layer
of
adsorbent of desired thickness over
the entire surface of the plate.
Procedure
—
[
NOTE
—
In this
procedure, use Purified Water that is obtained by
distillation.] Clean the glass plates
scrupulously, using an appropriate cleaning
solution (see
Cleaning Glass
Apparatus
1051
), rinsing
them with copious
quantities of water
until the water runs off the plates without
leaving any visible water
or oily
spots, then dry. It is important that the plates
be completely free from lint and
dust
when the adsorbent is applied.
Arrange the plate or plates on the
aligning tray, place a 5- ×
20-cm plate
adjacent to
the front edge of the first
square plate and another 5- ×
20-cm
plate adjacent to the
rear edge of the
last square, and secure all of the plates so that
they will not slip
during the
application of the adsorbent. Position the
spreader on the end plate
opposite the
raised end of the aligning tray. Mix 1 part of
adsorbent with 2 parts of
water (or in
the ratio suggested by the supplier) by shaking
vigorously for 30 seconds
in a glass-
stoppered conical flask, and transfer the slurry
to the spreader. Usually 30 g
of
adsorbent and 60 mL of water are sufficient for
five 20- ×
20-cm plates. Complete
the application of adsorbents using
plaster of Paris binder within 2 minutes of the
addition of the water, because
thereafter the mixture begins to harden. Draw the
spreader smoothly over the plates
toward the raised end of the aligning tray, and
remove the spreader when it is on the
end plate next to the raised end of the aligning
tray. (Wash away all traces of
adsorbent from the spreader immediately after
use.)
Allow the plates to remain
undisturbed for 5 minutes, then transfer the
square plates,
layer side up, to the
storage rack, and dry at 105
for 30 minutes. Preferably place the
rack at an angle in the drying oven to
prevent the condensation of moisture on the
back sides of plates in the rack. When
the plates are dry, allow them to cool to room
temperature, and inspect the uniformity
of the distribution and the texture of the
adsorbent layer; transmitted light will
show uniformity of distribution, and reflected
light
will show uniformity of texture.
Store the satisfactory plates over silica gel in a
suitable
chamber.
COLUMN CHROMATOGRAPHY
Apparatus
—
The
apparatus required for column chromatographic
procedures is
simple, consisting only
of the chromatographic tube itself and a tamping
rod, which
may be needed to pack a
pledget of glass wool or cotton, if needed, in the
base of the
tube and compress the
adsorbent or slurry uniformly within the tube. In
some cases a
porous glass disk is
sealed at the base of the tube in order to support
the contents.
The tube is cylindrical
and is made of glass, unless another material is
specified in the
individual monograph.
A smaller-diameter delivery tube is fused or
otherwise attached
by a leakproof joint
to the lower end of the main tube. Column
dimensions are variable;
the dimensions
of those commonly used in pharmaceutical analysis
range from 10 to
30 mm in uniform
inside diameter and 150 to 400 mm in length,
exclusive of the
delivery tube. The
delivery tube, usually 3 to 6 mm in inside
diameter, may include a
stopcock for
accurate control of the flow rate of solvents
through the column. The
tamping rod, a
cylindrical ram firmly attached to a shaft, may be
constructed of plastic,
glass,
stainless steel, or aluminum, unless another
material is specified in the
individual
monograph. The shaft of the rod is substantially
smaller in diameter than the
column and
is not less than 5 cm longer than the effective
length of the column. The
ram has a
diameter about 1 mm smaller than the inside
diameter of the column.
Column Adsorption Chromatography
The adsorbent (such as
activated alumina or silica gel, calcined
diatomaceous silica,
or chromatographic
purified siliceous earth) as a dry solid or as a
slurry is packed into
a glass or quartz
chromatographic tube. A solution of the drug in a
small amount of
solvent is added to the
top of the column and allowed to flow into the
adsorbent. The
drug principles are
quantitatively removed from the solution and are
adsorbed in a
narrow transverse band at
the top of the column. As additional solvent is
allowed to
flow through the column,
either by gravity or by application of air
pressure, each
substance progresses
down the column at a characteristic rate resulting
in a spatial
separation to give what is
known as the
chromatogram.
The rate of movement for a
given substance is affected by several
variables, including the adsorptive power of the
adsorbent and its particle size and
surface area; the nature and polarity of the
solvent;
the hydrostatic head or
applied pressure; and the temperature of the
chromatographic
system.
If
the separated compounds are colored or if they
fluoresce under UV light, the
adsorbent
column may be extruded and, by transverse cuts,
the appropriate
segments may then be
isolated. The desired compounds are then extracted
from
each segment with a suitable
solvent. If the compounds are colorless, they may
be
located by means of painting or
spraying the extruded column with color-forming
reagents. Chromatographed radioactive
substances may be located by means of
Geiger-Mü
ller detectors or
similar sensing and recording instruments. Clear
plastic
tubing made of a material such
as nylon, which is inert to most solvents and
transparent to short-wavelength UV
light, may be packed with adsorbent and used as
a chromatographic column. Such a column
may be sliced with a sharp knife without
removing the packing from the tubing.
If a fluorescent adsorbent is used, the column
may be marked under UV light in
preparation for slicing.
A ―flowing‖
chromatogram, which is extensively used, is
obtained by a procedure in
which
solvents are allowed to flow through the column
until the separated drug
appears in the
effluent solution, known as the ―eluate.‖ The drug
may be determined
in the eluate by
titration or by a spectrophotometric or
colorimetric method, or the
solvent may
be evaporated, leavi
ng the drug in more
or less pure form. If a second
drug
principle is involved, it is eluted by continuing
the first solvent or by passing a
solvent of stronger eluting power
through the column. The efficiency of the
separation
may be checked by obtaining
a thin-layer chromatogram on the individual
fractions.
A modified procedure for
adding the mixture to the column is sometimes
employed.
The drug, in a solid form,
and, as in the case of a powdered tablet, without
separation
from the excipients, is
mixed with some of the adsorbent and added to the
top of a
column. The subsequent flow of
solvent moves the drug down the column in the
manner described.
Column
Partition Chromatography
In
partition chromatography the substances to be
separated are partitioned between
two
immiscible liquids, one of which, the immobile
phase, is adsorbed on a
Solid
Support,
thereby presenting
a very large surface area to the flowing solvent
or mobile
phase. The exceedingly high
number of successive liquid-liquid contacts allows
an
efficiency of separation not
achieved in ordinary liquid-liquid extraction.
The
Solid Support
is usually polar, and the adsorbed
immobile phase more polar than
the
mobile phase. The
Solid Support
that is most widely used is
chromatographic
siliceous earth having
a particle size suitable to permit proper flow of
eluant.
1
In
reverse-phase partition chromatography
the adsorbed immobile phase is less polar
than the mobile phase and the solid
adsorbent is rendered nonpolar by treatment with
a silanizing agent, such as
dichlorodimethylsilane, to give silanized
chromatographic
siliceous earth.
The sample to be chromatographed is
usually introduced into the chromatographic
system in one of two ways: (a) a
solution of the sample in a small volume of the
mobile phase is added to the top of the
column; or, (b) a solution of the sample in a
small volume of the immobile phase is
mixed with the
Solid Support
and transferred to
the
column as a layer above a bed of a mixture of
immobile phase with adsorbent.
Development and elution are
accomplished with flowing solvent as before. The
mobile
solvent usually is saturated
with the immobile solvent before use.
In conventional liquid-liquid partition
chromatography, the degree of partition of a
given compound between the two liquid
phases is expressed by its partition or
distribution coefficient. In the case
of compounds that dissociate, distribution can be
controlled by modifying the pH,
dielectric constant, ionic strength, and other
properties
of the two phases. Selective
elution of the components of a mixture can be
achieved
by successively changing the
mobile phase to one that provides a more favorable
partition coefficient, or by changing
the pH of the immobile phase
in situ
with a mobile
phase
consisting of a solution of an appropriate acid or
base in an organic solvent.
Unless
otherwise specified in the individual monograph,
assays and tests that employ
column
partition chromatography are performed according
to the following general
methods.
Solid Support
—
Use purified siliceous earth. Use silanized
chromatographic siliceous
earth for
reverse-phase partition chromatography.
Stationary
Phase
—
Use the solvent or
solution specified in the individual monograph.
If a mixture of liquids is to be used
as the
Stationary Phase,
mix
them prior to the
introduction of the
Solid Support.
Mobile Phase
—
Use
the solvent or solution specified in the
individual monograph.
Equilibrate it
with water if the
Stationary Phase
is an aqueous solution; if the
Stationary Phase
is a polar
organic fluid, equilibrate with that fluid.
Preparation of
Chromatographic Column
—
Unless otherwise specified in the
individual monograph, the
chromatographic tube is about 22 mm in inside
diameter
and 200 to 300 mm in length,
without porous glass disk, to which is attached a
delivery tube, without stopcock, about
4 mm in inside diameter and about 50 mm in
length. Pack a pledget of fine glass
wool in the base of the tube. Place the specified
volume of
Stationary
Phase
in a 100- to 250-mL beaker, add
the specified amount of
Solid Support,
and mix to produce a homogeneous,
fluffy mixture. Transfer this
mixture
to the chromatographic tube, and tamp, using
gentle pressure, to obtain a
uniform
mass. If the specified amount of
Solid
Support
is more than 3 g, transfer the
mixture to the column in portions of
approximately 2 g, and tamp each portion. If the
assay or test requires a multisegment
column, with a different
Stationary
Phase
specified for each segment, tamp
after the addition of each segment, and add each
succeeding segment directly to the
previous one.
If a solution
of the analyte is incorporated in the
Stationary Phase,
complete
the
quantitative transfer to the
chromatographic tube by scrubbing the beaker used
for the
preparation of the test mixture
with a mixture of about 1 g of
Solid
Support
and several
drops of
the solvent used to prepare the test solution.
Pack a pledget of fine glass wool above
the completed column packing. The
Mobile
Phase
flows through a properly packed column
as a moderate stream or, if
reverse-
phase chromatography is applied, as a slow
trickle.
Procedure
—
Transfer the
Mobile Phase
to
the column space above the column
packing, and allow it to flow through
the column under the influence of gravity. Rinse
the tip of the chromatographic column
with about 1 mL of
Mobile Phase
before each
change in
composition of
Mobile Phase
and after completion of the elution. If
the
analyte is introduced into the
column as a solution in the
Mobile
Phase,
allow it to
pass
completely into the column packing, then add
Mobile Phase
in several
small
portions, allowing each to drain
completely, before adding the bulk of the
Mobile
Phase.
Where the assay or test requires the
use of multiple chromatographic columns
mounted in series and the addition of
Mobile Phase
in divided
portions is specified,
allow each
portion to drain completely through each column,
and rinse the tip of each
with
Mobile Phase
prior to the
addition of each succeeding portion.
GAS
CHROMATOGRAPHY
The distinguishing features of gas
chromatography are a gaseous mobile phase and a
solid or immobilized liquid stationary
phase. Liquid stationary phases are available in
packed or capillary columns. In the
packed columns, the liquid phase is deposited on
a finely divided, inert solid support,
such as diatomaceous earth, porous polymer, or
graphitized carbon, which is packed
into a column that is typically 2 to 4 mm in
internal
diameter and 1 to 3 m in
length. In capillary columns, which contain no
packing, the
liquid phase is deposited
on the inner surface of the column and may be
chemically
bonded to it. In gas-solid
chromatography, the solid phase is an active
adsorbent,
such as alumina, silica, or
carbon, packed into a column. Polyaromatic porous
resins,
which are sometimes used in
packed columns, are not coated with a liquid
phase.
When a vaporized compound is
introduced into the carrier gas and carried into
the
column, it is partitioned between
the gas and stationary phases by a dynamic
countercurrent distribution process.
The compound is carried down the column by the
carrier gas, retarded to a greater or
lesser extent by sorption and desorption on the
stationary phase. The elution of the
compound is characterized by the partition ratio,
k
?
,
a
dimensionless quantity also called the capacity
factor (see
Glossary of Symbols
for the definition of symbols). It is
equivalent to the ratio of the time required for
the
compound to flow through the column
(the retention time) to the elution time of an
unretained compound. The value of the
capacity factor depends on the chemical
nature of the compound, the nature,
amount, and surface area of the liquid phase, the
column temperature, and the gas flow
rate. Under a specified set of experimental
conditions, a characteristic capacity
factor exists for every compound. Separation by
gas chromatography occurs only if the
compounds concerned have different capacity
factors.
Apparatus
—
A gas
chromatograph consists of a carrier gas source, an
injection port,
column, detector, and
recording device. The injection port, column, and
detector are
temperature-controlled.
The typical carrier gas is helium, nitrogen, or
hydrogen,
depending on the column and
detector in use. The gas is supplied from a
high-pressure cylinder or high-purity
gas generator and passes through suitable
pressure-reducing valves and a flow
meter to the injection port and column.
Compounds to be chromatographed, either
in solution or as gases, are injec
ted
into
the gas stream at the injection
port. Depending upon the configuration of the
apparatus, the test mixture may be
injected directly into the column or be vaporized
in
the injection port and mixed into
the flowing carrier gas prior to entering the
column.
Once in the column,
compounds in the test mixture are separated by
virtue of
differences in their capacity
factors, which in turn depend upon vapor pressure
and
degree of interaction with the
stationary phase. The capacity factor, which
governs
resolution, retention times,
and column efficiencies of components of the test
mixture,
is also temperature-dependent.
The use of temperature-programmable column ovens
takes advantage of this dependence to
achieve efficient separation of compounds
differing widely in vapor pressure.
As resolved compounds emerge separately
from the column, they pass through a
differential detector, which responds
to the amount of each compound present. The
type of detector to be used depends
upon the nature of the compounds to be
analyzed and is specified in the
individual monograph. Detectors are heated to
prevent condensation of the eluting
compounds.
Detector output is recorded
as a function of time, producing a chromatogram,
which
consists of a series of peaks on
a time axis. Each peak represents a compound in
the
vaporized test mixture, although
some peaks may overlap. The elution time is a
characteristic of an individual
compound; and the instrument response, measured as
peak area or peak height, is a function
of the amount present.
Injectors
—
Sample
injection devices range from simple syringes to
fully programmable
automatic injectors.
The amount of sample that can be injected into a
capillary column
without overloading is
small compared to the amount that can be injected
into packed
columns, and may be less
than the smallest amount that can be manipulated
satisfactorily by syringe. Capillary
columns, therefore, often are used with injectors
able to split samples into two
fractions, a small one that enters the column and
a large
one that goes to waste. Such
injectors may be used in a
splitless
mode
for analyses of
trace
or minor components.
Purge and trap
injectors are equipped with a sparging device by
which volatile
compounds in solution
are carried into a low-temperature trap. When
sparging is
complete, trapped compounds
are desorbed into the carrier gas by rapid heating
of
the temperature-programmable trap.
Headspace injectors are equipped with a
thermostatically controlled sample heating
chamber. Solid or liquid samples in
tightly closed containers are heated in the
chamber for a fixed period of time,
allowing the volatile components in the sample to
reach an equilibrium between the
nongaseous phase and the gaseous or headspace
phase.
After this
equilibrium has been established, the injector
automatically introduces a
fixed amount
of the headspace in the sample container into the
gas chromatograph.
Columns
—
Capillary
columns, which are usually made of fused silica,
are typically 0.2
to 0.53 mm in
internal diameter and 5 to 60 m in length. The
liquid or stationary phase,
which is
sometimes chemically bonded to the inner surface,
is 0.1 to 1.0 ?
m thick,
although nonpolar stationary phases may
be up to 5 ?
m thick. A list of liquid
phases in
current use is given in the
section
Chromatographic
Reagents.
Packed
columns, made of glass or metal, are 1 to 3 m in
length with internal
diameters of 2 to
4 mm. Those used for analysis typically are porous
polymers or solid
supports with liquid
phase loadings of about 5% (w/w). High-capacity
columns, with
liquid phase loadings of
about 20% (w/w), are used for large test specimens
and for
the determination of low
molecular weight compounds such as water. The
capacity
required influences the choice
of solid support.
Supports for analysis
of polar compounds on low-capacity, low-polarity
liquid phase
columns must be inert to
avoid peak tailing. The reactivity of support
materials can be
reduced by silanizing
prior to coating with liquid phase. Acid-washed,
flux-calcined
diatomaceous earth is
often used for drug analysis. Support materials
are available in
various mesh sizes,
with 80- to 100-mesh and 100- to 120-mesh being
most
commonly used with 2- to 4-mm
columns. Supports and liquid phases are listed in
the
section
Chromatographic
Reagents.
Retention time and the peak efficiency
depend on the carrier gas flow rate; retention
time is also directly proportional to
column length, while resolution is proportional to
the square root of the column length.
For packed columns, the carrier gas flow rate is
usually expressed in mL per minute at
atmospheric pressure and room temperature. It
is measured at the detector outlet with
a flowmeter while the column is at operating
temperature. The linear flow rate
through a packed column is inversely proportional
to
the square of the column diameter
for a given flow volume. Flow rates of 60 mL per
minute in a 4-mm column and 15 mL per
minute in a 2-mm column give identical
linear flow rates and thus similar
retention times. Unless otherwise specified in the
individual monograph, flow rates for
packed columns are about 30 to 60 mL per
minute. For capillary columns, linear
flow velocity is often used instead of flow rate.
This is conveniently determined from
the length of the column and the retention time
of a dilute methane sample, provided a
flame-ionization detector is in use. At high
operating temperatures there is
sufficient vapor pressure to result in a gradual
loss of
liquid phase, a process called
bleeding.
-
-
-
-
-
-
-
-
-
上一篇:以万字开头的成语大全及解释
下一篇:花开头的成语大全