关键词不能为空

当前您在: 主页 > 英语 >

(621)CHROMATOGRAPHY《色谱法》USP31

作者:高考题库网
来源:https://www.bjmy2z.cn/gaokao
2021-02-11 17:27
tags:

-

2021年2月11日发(作者:grossweight)


621



CHROMATOGRAPHY





INTRODUCTION




This chapter defines the terms and procedures used in chromatography and provides


general information. Specific requirements for chromatographic procedures for drug


substances and dosage forms, including adsorbent and developing solvents, are


given in the individual monographs.


Chromatography is defined as a procedure by which solutes are separated by a


dynamic differential migration process in a system consisting of two or more phases,


one of which moves continuously in a given direction and in which the individual


substances exhibit different mobilities by reason of differences in adsorption, partition,


solubility, vapor pressure, molecular size, or ionic charge density. The individual


substances thus separated can be identified or determined by analytical procedures.


The general chromatographic technique requires that a solute undergo distribution


between two phases, one of them fixed (stationary phase), the other moving (mobile


phase). It is the mobile phase that transfers the solute through the medium until it


eventually emerges separated from other solutes that are eluted earlier or later.


Generally, the solute is transported through the separation medium by means of a


flowing stream of a liquid or a gaseous solvent known as the ―eluant.‖ The stationary


phase may act through adsorption, as in the case of adsorbents such as activated


alumina and silica gel, or it may act by dissolving the solute, thus partitioning the latter


between the stationary and mobile phases. In the latter process, a liquid coated onto


an inert support, or chemically bonded onto silica gel, or directly onto the wall of a


fused silica capillary, serves as the stationary phase. Partitioning is the predominant


mechanism of separation in gas



liquid chromatography, paper chromatography, in


forms of column chromatography, and in thin-layer chromatography designated as


liquid- liquid chromatography. In practice, separations frequently result from a


combination of adsorption and partitioning effects. Other separation principles include


ion exchange, ion-pair formation, size exclusion, hydrophobic interaction, and chiral


recognition.


The types of chromatography useful in qualitative and quantitative analysis that are


employed in the


USP


procedures are column, gas, paper, thin-layer, (including


high-performance thin-layer chromatography), and pressurized liquid chromatography


(commonly called high- pressure or high-performance liquid chromatography). Paper


and thin-layer chromatography are ordinarily more useful for purposes of identification,


because of their convenience and simplicity. Column chromatography offers a wider


choice of stationary phases and is useful for the separation of individual compounds,


in quantity, from mixtures. Modern high-performance thin-layer chromatography, gas


chromatography, and pressurized liquid chromatography require more elaborate


apparatus but usually provide high resolution and identify and quantitate very small


amounts of material.


Use of Reference Substances in Identity Tests



In paper and thin- layer


chromatography, the ratio of the distance (this distance being measured to the point of


maximum intensity of the spot or zone) traveled on the medium by a given compound


to the distance traveled by the front of the mobile phase, from the point of application


of the test substance, is designated as the


R


F


value of the compound. The ratio


between the distances traveled by a given compound and a reference substance is


the


R


R


value.


R


F


values vary with the experimental conditions, and thus identification is


best accomplished where an authentic specimen of the compound in question is used


as a reference substance on the same chromatogram.



For this purpose, chromatograms are prepared by applying on the thin-layer


adsorbent or on the paper in a straight line, parallel to the edge of the


chromatographic plate or paper, solutions of the substance to be identified, the


authentic specimen, and a mixture of nearly equal amounts of the substance to be


identified and the authentic specimen. Each sample application contains


approximately the same quantity by weight of material to be chromatographed. If the


substance to be identified and the authentic specimen are identical, all


chromatograms agree in color and


R


F


value and the mixed chromatogram yields a


single spot; i.e.,


R


R


is 1.0.


Location and Identification of Components



The spots produced by paper or


thin-layer chromatography may be located by: (1) direct inspection if the compounds


are visible under white or either short-wavelength (254 nm) or long-wavelength (360


nm) UV light, (2) inspection in white or UV light after treatment with reagents that will


make the spots visible (reagents are most conveniently applied with an atomizer), (3)


use of a Geiger- Mü


ller counter or autoradiographic techniques in the case of the


presence of radioactive substances, or (4) evidence resulting from stimulation or


inhibition of bacterial growth by the placing of removed portions of the adsorbent and


substance on inoculated media.



In open-column chromatography, in pressurized liquid chromatography performed


under conditions of constant flow rate, and in gas chromatography, the retention time,


t,


defined as the time elapsed between sample injection and appearance of the peak


concentration of the eluted sample zone, may be used as a parameter of identification.


Solutions of the substance to be identified or derivatives thereof, of the reference


compound, and of a mixture of equal amounts of these two are chromatographed


successively on the same column under the same chromatographic conditions. Only


one peak should be observed for the mixture. The ratio of the retention times of the


test substance, the reference compound, and a mixture of these, to the retention time


of an internal standard is called the relative retention time


R


R


and is also used


frequently as a parameter of identification.


The deviations of


R


R


,


R


F


, or


t


values measured for the test substance from the values


obtained for the reference compound and mixture should not exceed the reliability


estimates determined statistically from replicate assays of the reference compound.


Chromatographic identification by these methods under given conditions strongly


indicates identity but does not constitute definitive identification. Coincidence of


identity parameters under three to six different sets of chromatographic conditions


(


temperatures, column packings, adsorbents, eluants, developing solvents, various


chemical derivatives, etc


.) increases the probability that the test and reference


substances are identical. However, many isomeric compounds cannot be separated.


Specific and pertinent chemical, spectroscopic, or physicochemical identification of


the eluted component combined with chromatographic identity is the most valid


criterion of identification. For this purpose, the individual components separated by


chromatography may be collected for further identification.



PAPER CHROMATOGRAPHY




In paper chromatography the adsorbent is a sheet of paper of suitable texture and


thickness. Chromatographic separation may proceed through the action of a single


liquid phase in a process analogous to adsorption chromatography in columns. Since


the natural water content of the paper, or selective imbibition of a hydrophilic


component of the liquid phase by the paper fibers, may be regarded as a stationary


phase, a partitioning mechanism may contribute significantly to the separation.


Alternatively, a two-phase system may be used. The paper is impregnated with one of


the phases, which then remains stationary (usually the more polar phase in the case


of unmodified paper). The chromatogram is developed by slow passage of the other,


mobile phase over the sheet. Development may be ascending, in which case the


solvent is carried up the paper by capillary forces, or descending, in which case the


solvent flow is also assisted by gravitational force.


Differences in the value of


R


F


have been reported where chromatograms developed in


the direction of the paper grain (machine direction) are compared with others


developed at right angles to the grain; therefore, the orientation of paper grain with


respect to solvent flow should be maintained constant in a series of chromatograms.


(The machine direction is usually designated by the manufacturer on packages of


chromatography paper.)


Descending Chromatography



In descending chromatography, the mobile phase flows downward on the


chromatographic sheet.


Apparatus



The essential equipment for descending chromatography consists of the


following:



A


vapor-tight chamber


provided with inlets for addition of solvent or for releasing


internal pressure. The chamber is constructed preferably of glass, stainless steel, or


porcelain and is so designed as to permit observation of the progress of the


chromatographic run without opening of the chamber. Tall glass cylinders are


convenient if they are made vapor-tight with suitable covers and a sealing compound.


A


rack of corrosion-resistant material


about 5 cm shorter than the inside height of the


chamber. The rack serves as a support for solvent troughs and for antisiphon rods


which, in turn, hold up the chromatographic sheets.


One or more


glass troughs


capable of holding a volume of solvent greater than that


needed for one chromatographic run. The troughs must also be longer than the width


of the chromatographic sheets.


Heavy glass antisiphon rods


to be supported by the rack and running outside of,


parallel to, and slightly above the edge of the glass trough.


Chromatographic sheets


of special filter paper at least 2.5 cm wide and not wider than


the length of the troughs are cut to a length approximately equal to the height of the


chamber. A fine pencil line is drawn horizontally across the filter paper at a distance


from one end such that, when the sheet is suspended from the antisiphon rods with


the upper end of the paper resting in the trough and the lower portion hanging free


into the chamber, the line is located a few centimeters below the rods. Care is


necessary to avoid contaminating the filter paper by excessive handling or by contact


with dirty surfaces.


Procedure



The substance or substances to be analyzed are dissolved in a suitable


solvent. Convenient volumes, delivered from suitable micropipets, of the resulting


solution, normally containing 1 to 20 ?


g of the compound, are placed in 6- to 10-mm


spots not less than 3 cm apart along the pencil line. If the total volume to be applied


would produce spots of a diameter greater than 6 to 10 mm, it is applied in separate


portions to the same spot, each portion being allowed to dry before the next is added.



The spotted chromatographic sheet is suspended in the chamber by use of the


antisiphon rod, which holds the upper end of the sheet in the solvent trough. The


bottom of the chamber is covered with the prescribed solvent system. Saturation of


the chamber with solvent vapor is facilitated by lining the inside walls with paper that is


wetted with the prescribed solvent system. It is important to ensure that the portion of


the sheet hanging below the rods is freely suspended in the chamber without touching


the rack or the chamber walls or the fluid in the chamber. The chamber is sealed to


allow equilibration (saturation) of the chamber and the paper with the solvent vapor.


Any excess pressure is released as necessary. For large chambers, equilibration


overnight may be necessary.


A volume of the mobile phase in excess of the volume required for complete


development of the chromatogram is saturated with the immobile phase by shaking.


After equilibration of the chamber, the prepared mobile solvent is introduced into the


trough through the inlet. The inlet is closed and the mobile solvent phase is allowed to


travel the desired distance down the paper. Precautions must be taken against


allowing the solvent to run down the sheet when opening the chamber and removing


the chromatogram. The location of the solvent front is quickly marked, and the sheets


are dried.


The chromatogram is observed and measured directly or after suitable development


to reveal the location of the spots of the isolated drug or drugs. The paper section(s)


predetermined to contain the isolated drug(s) may be cut out and eluted by an


appropriate solvent, and the solutions may be made up to a known volume and


quantitatively analyzed by appropriate chemical or instrumental techniques. Similar


procedures should be conducted with various amounts of similarly spotted reference


standard on the same paper in the concentration range appropriate to prepare a valid


calibration curve.


Ascending Chromatography



In ascending chromatography, the lower edge of the sheet (or strip) is dipped into the


mobile phase to permit the mobile phase to rise on the chromatographic sheet by


capillary action.


Apparatus



The essential equipment for ascending chromatography is substantially


the same as that described under


Descending Chromatography.




Procedure



The test materials are applied to the chromatographic sheets as directed


under


Descending Chromatography,


and above the level to which the paper is dipped


into the developing solvent. The bottom of the developing chamber is covered with the


developing solvent system. If a two-phase system is used, both phases are added. It


is also desirable to line the walls of the chamber with paper and to saturate this lining


with the solvent system. Empty solvent troughs are placed on the bottom of the


chamber, and the chromatographic sheets are suspended so that the end on which


the spots have been added hangs free inside the empty trough.



The chamber is sealed, and equilibration is allowed to proceed as described under


Descending Chromatography.


Then the developing solvent (mobile phase) is added


through the inlet to the trough in excess of the solvent required for complete


moistening of the chromatographic sheet. The chamber is resealed. When the solvent


front has reached the desired height, the chamber is opened and the sheet is


removed and dried.


Quantitative analyses of the spots may be conducted as described under


Descending


Chromatography.





THIN-LAYER CHROMATOGRAPHY




In thin-layer chromatography, the adsorbent is a relatively thin, uniform layer of dry,


finely powdered material applied to a glass, plastic, or metal sheet or plate, glass


plates



being most commonly employed. The coated plate can be considered an ―open


chromatographic column‖ and the separations achieved may be based upon


adsorption, partition, or a combination of both effects, depending on the particular


type of stationary phase, its preparation, and its use with different solvents. Thin-layer


chromatography on ion- exchange layers can be used for the fractionation of polar


compounds. Presumptive identification can be effected by observation of spots or


zones of identical


R


F


value and about equal magnitude obtained, respectively, with an


unknown and a reference sample chromatographed on the same plate. A visual


comparison of the size or intensity of the spots or zones may serve for


semiquantitative estimation. Quantitative measurements are possible by means of


densitometry (absorbance or fluorescence measurements), or the spots may be


carefully removed from the plate, followed by elution with a suitable solvent and


spectrophotometric measurement. For two-dimensional thin-layer chromatography,


the chromatographed plate is turned at a right angle and again chromatographed,


usually in another chamber equilibrated with a different solvent system.


Apparatus



Acceptable apparatus and materials for thin-layer chromatography


consist of the following.



A


TLC or HPTLC plate.


The chromatography is generally carried out using


precoated


plates


or


sheets


(on glass, aluminum, or polyester support) of suitable size. It may be


necessary to clean the plates prior to separation. This can be done by migration of, or


immersion in, an appropriate solvent. The plates may also be impregnated by


procedures such as development, immersion, or spraying. At the time of use, the


plates may be activated, if necessary, by heating in an oven at 120



for 20 minutes.


The


stationary phase


of TLC plates has an average particle size of 10



15 ?


m, and


that of HPTLC plates an average particle size of 5 ?


m. Commercial plates with a


preadsorbant zone can be used if they are specified in a monograph. Sample applied


to the preabsorbant region develops into sharp, narrow bands at the


preabsorbant-sorbent interface. Alternatively, flat


glass plates


of convenient size,


typically 20 cm ×


20 cm can be coated as described under


Preparation of


Chromatographic Plates.




A suitable


manual, semiautomatic, or automatic application device


can be used to


ensure proper positioning of the plate and proper transfer of the sample, with respect


to volume and position, onto the plate. Alternatively, a


template


can be used to guide


in manually placing the test spots at definite intervals, to mark distances as needed,


and to aid in labeling the plates. For the proper application of the solutions,


micropipets, microsyringes, or calibrated disposable capillaries


are recommended.


For ascending development, a


chromatographic chamber


made of inert, transparent


material and having the following specifications is used: a flat bottom or twin trough, a


tightly fitted lid, and a size suitable for the plates. For horizontal development, the


chamber is provided with a reservoir for the mobile phase, and it also contains a


device for directing the mobile phase to the stationary phase.


Devices for transfer of reagents


onto the plate by spraying, immersion, or exposure to


vapor and devices to facilitate any necessary heating for visualization of the


separated spots or zones.


A


UV light source


suitable for observations under short (254 nm) and long (365 nm)


wavelength UV light.


A suitable


device for documentation


of the visualized chromatographic result.


Procedure



Apply the prescribed volume of the test solution and the standard


solution in sufficiently small portions to obtain circular spots of 2 to 5 mm in diameter


(1 to 2 mm on HPTLC plates) or bands of 10 to 20 mm by 1 to 2 mm (5 to 10 mm by


0.5 to 1 mm on HPTLC plates) at an appropriate distance from the lower


edge



during chromatography the application position must be at least 3 mm (HPTLC)


or 5 mm (TLC) above the level of the developing solvent



and from the sides of the


plate. Apply the solutions on a line parallel to the lower edge of the plate with an


interval of at least 10 mm (5 mm on HPTLC plates) between the centers of spots


or 4


mm (2 mm on HPTLC plates) between the edges of bands, and allow to dry.



Ascending Development



Line at least one wall of the chromatographic chamber


with filter paper. Pour into the chromatographic chamber a quantity of the mobile


phase sufficient for the size of the chamber to give, after impregnation of the filter


paper, a level of depth appropriate to the dimension of the plate used. For saturation


of the chromatographic chamber, close the lid, and allow the system to equilibrate.


Unless otherwise indicated, the chromatographic separation is performed in a


saturated chamber.



Place the plate in the chamber, ensuring that the plate is as vertical as possible and


that the spots or bands are above the surface of the mobile phase, and close the


chamber. The stationary phase faces the inside of the chamber. Remove the plate


when the mobile phase has moved over the prescribed distance. Dry the plate, and


visualize the chromatograms as prescribed. For two-dimensional chromatography,


dry the plates after the first development, and carry out a second development in a


direction perpendicular to that of the first development.


Horizontal Development



Introduce a sufficient quantity of the developing solvent


into the reservoir of the chamber using a syringe or pipet. Place the plate horizontally


in the chamber, connect the mobile phase direction device according to the


manufacturer's instructions, and close the chamber. If prescribed, develop the plate


starting simultaneously at both ends. Remove the plate when the mobile phase has


moved over the distance prescribed in the monograph. Dry the plate, and visualize the


chromatograms as prescribed.



For two-dimensional chromatography, dry the plates after the first development, and


carry out a second development in a direction perpendicular to that of the first


development.


Detection



Observe the dry plate first under short-wavelength UV light (254 nm) and


then under long- wavelength UV light (365 nm) or as stated in the monograph. If


further directed, spray, immerse, or expose the plate to vapors of the specified


reagent, heat the plate when required, observe, and compare the test chromatogram


with the standard chromatogram. Document the plate after each observation.


Measure and record the distance of each spot or zone from the point of origin, and


indicate for each spot or zone the wavelength under which it was observed.


Determine the


R


F


values for the principal spots or zones (see


Glossary of Symbols


).



Quantitative Measurement



Using appropriate instrumentation, substances


separated by TLC and responding to ultraviolet-visible (UV-Vis) irradiation prior to or


after derivatization can be determined directly on the plate. While moving the plate or


the measuring device, the plate is examined by measuring the reflectance of the


incident light. Similarly, fluorescence may be measured using an appropriate optical


system. Substances containing radionuclides can be quantified in three ways: (1)


directly by moving the plate alongside a suitable counter or vice versa; (2) by cutting


the plates into strips and measuring the radioactivity on each individual strip using a


suitable counter; or (3) by scraping off the stationary phase, dissolving it in a suitable


scintillation cocktail, and measuring the radioactivity using a liquid scintillation counter


(see


Radioactivity


821


).



The apparatus for direct quantitative measurement on the plate is a densitometer that


is composed of a mechanical device to move the plate or the measuring device along


the


x


-axis and the


y


-axis, a recorder, a suitable integrator or a computer; and, for


substances responding to UV-Vis irradiation, a photometer with a source of light, an


optical device capable of generating monochromatic light, and a photo cell of


adequate sensitivity, all of which are used for the measurement of reflectance. In the


case where fluorescence is measured, a suitable filter is also required to prevent the


light used for excitation from reaching the photo cell while permitting the emitted light


or specific portions thereof to pass. The linearity range of the counting device must be


verified.


For quantitative tests, it is necessary to apply to the plate not fewer than three


standard solutions of the substance to be examined, the concentrations of which span


the expected value in the test solution (e.g., 80%, 100%, and 120%). Derivatize with


the prescribed reagent, if necessary, and record the reflectance or fluorescence in the


chromatograms obtained. Use the measured results for the calculation of the amount


of substance in the test solution.


Preparation of Chromatographic Plates





Apparatus





Flat


glass plates


of convenient size, typically 20 cm ×


20 cm.


An


aligning tray


or a flat surface upon which to align and rest the plates during the


application of the adsorbent.


A


storage rack


to hold the prepared plates during drying and transportation. The rack


holding the plates should be kept in a desiccator or be capable of being sealed in


order to protect the plates from the environment after removal from the drying oven.


The


adsorbent


consists of finely divided adsorbent materials, normally 5 to 40 ?


m in


diameter, suitable for chromatography. It can be applied directly to the glass plate or


can be bonded to the plate by means of plaster of Paris [calcium sulfate hemihydrate


(at a ratio of 5% to 15%)] or with starch paste or other binders. The plaster of Paris will


not yield as hard a surface as will the starch, but it is not affected by strongly oxidizing


spray reagents. The adsorbent may contain fluorescing material to aid in the


visualization of spots that absorb UV light.


A


spreader,


which, when moved over the glass plate, will apply a uniform layer of


adsorbent of desired thickness over the entire surface of the plate.


Procedure



[


NOTE



In this procedure, use Purified Water that is obtained by


distillation.] Clean the glass plates scrupulously, using an appropriate cleaning


solution (see


Cleaning Glass Apparatus


1051


), rinsing them with copious


quantities of water until the water runs off the plates without leaving any visible water


or oily spots, then dry. It is important that the plates be completely free from lint and


dust when the adsorbent is applied.



Arrange the plate or plates on the aligning tray, place a 5- ×


20-cm plate adjacent to


the front edge of the first square plate and another 5- ×


20-cm plate adjacent to the


rear edge of the last square, and secure all of the plates so that they will not slip


during the application of the adsorbent. Position the spreader on the end plate


opposite the raised end of the aligning tray. Mix 1 part of adsorbent with 2 parts of


water (or in the ratio suggested by the supplier) by shaking vigorously for 30 seconds


in a glass- stoppered conical flask, and transfer the slurry to the spreader. Usually 30 g


of adsorbent and 60 mL of water are sufficient for five 20- ×


20-cm plates. Complete


the application of adsorbents using plaster of Paris binder within 2 minutes of the


addition of the water, because thereafter the mixture begins to harden. Draw the


spreader smoothly over the plates toward the raised end of the aligning tray, and


remove the spreader when it is on the end plate next to the raised end of the aligning


tray. (Wash away all traces of adsorbent from the spreader immediately after use.)


Allow the plates to remain undisturbed for 5 minutes, then transfer the square plates,


layer side up, to the storage rack, and dry at 105



for 30 minutes. Preferably place the


rack at an angle in the drying oven to prevent the condensation of moisture on the


back sides of plates in the rack. When the plates are dry, allow them to cool to room


temperature, and inspect the uniformity of the distribution and the texture of the


adsorbent layer; transmitted light will show uniformity of distribution, and reflected light


will show uniformity of texture. Store the satisfactory plates over silica gel in a suitable


chamber.



COLUMN CHROMATOGRAPHY



Apparatus



The apparatus required for column chromatographic procedures is


simple, consisting only of the chromatographic tube itself and a tamping rod, which


may be needed to pack a pledget of glass wool or cotton, if needed, in the base of the


tube and compress the adsorbent or slurry uniformly within the tube. In some cases a


porous glass disk is sealed at the base of the tube in order to support the contents.


The tube is cylindrical and is made of glass, unless another material is specified in the


individual monograph. A smaller-diameter delivery tube is fused or otherwise attached


by a leakproof joint to the lower end of the main tube. Column dimensions are variable;


the dimensions of those commonly used in pharmaceutical analysis range from 10 to


30 mm in uniform inside diameter and 150 to 400 mm in length, exclusive of the


delivery tube. The delivery tube, usually 3 to 6 mm in inside diameter, may include a


stopcock for accurate control of the flow rate of solvents through the column. The


tamping rod, a cylindrical ram firmly attached to a shaft, may be constructed of plastic,


glass, stainless steel, or aluminum, unless another material is specified in the


individual monograph. The shaft of the rod is substantially smaller in diameter than the


column and is not less than 5 cm longer than the effective length of the column. The


ram has a diameter about 1 mm smaller than the inside diameter of the column.



Column Adsorption Chromatography



The adsorbent (such as activated alumina or silica gel, calcined diatomaceous silica,


or chromatographic purified siliceous earth) as a dry solid or as a slurry is packed into


a glass or quartz chromatographic tube. A solution of the drug in a small amount of


solvent is added to the top of the column and allowed to flow into the adsorbent. The


drug principles are quantitatively removed from the solution and are adsorbed in a


narrow transverse band at the top of the column. As additional solvent is allowed to


flow through the column, either by gravity or by application of air pressure, each


substance progresses down the column at a characteristic rate resulting in a spatial


separation to give what is known as the


chromatogram.


The rate of movement for a


given substance is affected by several variables, including the adsorptive power of the


adsorbent and its particle size and surface area; the nature and polarity of the solvent;


the hydrostatic head or applied pressure; and the temperature of the chromatographic


system.


If the separated compounds are colored or if they fluoresce under UV light, the


adsorbent column may be extruded and, by transverse cuts, the appropriate


segments may then be isolated. The desired compounds are then extracted from


each segment with a suitable solvent. If the compounds are colorless, they may be


located by means of painting or spraying the extruded column with color-forming


reagents. Chromatographed radioactive substances may be located by means of


Geiger-Mü


ller detectors or similar sensing and recording instruments. Clear plastic


tubing made of a material such as nylon, which is inert to most solvents and


transparent to short-wavelength UV light, may be packed with adsorbent and used as


a chromatographic column. Such a column may be sliced with a sharp knife without


removing the packing from the tubing. If a fluorescent adsorbent is used, the column


may be marked under UV light in preparation for slicing.


A ―flowing‖ chromatogram, which is extensively used, is obtained by a procedure in


which solvents are allowed to flow through the column until the separated drug


appears in the effluent solution, known as the ―eluate.‖ The drug may be determined


in the eluate by titration or by a spectrophotometric or colorimetric method, or the


solvent may be evaporated, leavi


ng the drug in more or less pure form. If a second


drug principle is involved, it is eluted by continuing the first solvent or by passing a


solvent of stronger eluting power through the column. The efficiency of the separation


may be checked by obtaining a thin-layer chromatogram on the individual fractions.


A modified procedure for adding the mixture to the column is sometimes employed.


The drug, in a solid form, and, as in the case of a powdered tablet, without separation


from the excipients, is mixed with some of the adsorbent and added to the top of a


column. The subsequent flow of solvent moves the drug down the column in the


manner described.


Column Partition Chromatography



In partition chromatography the substances to be separated are partitioned between


two immiscible liquids, one of which, the immobile phase, is adsorbed on a


Solid


Support,


thereby presenting a very large surface area to the flowing solvent or mobile


phase. The exceedingly high number of successive liquid-liquid contacts allows an


efficiency of separation not achieved in ordinary liquid-liquid extraction.


The


Solid Support


is usually polar, and the adsorbed immobile phase more polar than


the mobile phase. The


Solid Support


that is most widely used is chromatographic


siliceous earth having a particle size suitable to permit proper flow of eluant.


1


In


reverse-phase partition chromatography the adsorbed immobile phase is less polar


than the mobile phase and the solid adsorbent is rendered nonpolar by treatment with


a silanizing agent, such as dichlorodimethylsilane, to give silanized chromatographic


siliceous earth.


The sample to be chromatographed is usually introduced into the chromatographic


system in one of two ways: (a) a solution of the sample in a small volume of the


mobile phase is added to the top of the column; or, (b) a solution of the sample in a


small volume of the immobile phase is mixed with the


Solid Support


and transferred to


the column as a layer above a bed of a mixture of immobile phase with adsorbent.


Development and elution are accomplished with flowing solvent as before. The mobile


solvent usually is saturated with the immobile solvent before use.


In conventional liquid-liquid partition chromatography, the degree of partition of a


given compound between the two liquid phases is expressed by its partition or


distribution coefficient. In the case of compounds that dissociate, distribution can be


controlled by modifying the pH, dielectric constant, ionic strength, and other properties


of the two phases. Selective elution of the components of a mixture can be achieved


by successively changing the mobile phase to one that provides a more favorable


partition coefficient, or by changing the pH of the immobile phase


in situ


with a mobile


phase consisting of a solution of an appropriate acid or base in an organic solvent.


Unless otherwise specified in the individual monograph, assays and tests that employ


column partition chromatography are performed according to the following general


methods.


Solid Support



Use purified siliceous earth. Use silanized chromatographic siliceous


earth for reverse-phase partition chromatography.



Stationary Phase



Use the solvent or solution specified in the individual monograph.


If a mixture of liquids is to be used as the


Stationary Phase,


mix them prior to the


introduction of the


Solid Support.




Mobile Phase



Use the solvent or solution specified in the individual monograph.


Equilibrate it with water if the


Stationary Phase


is an aqueous solution; if the


Stationary Phase


is a polar organic fluid, equilibrate with that fluid.



Preparation of Chromatographic Column



Unless otherwise specified in the


individual monograph, the chromatographic tube is about 22 mm in inside diameter


and 200 to 300 mm in length, without porous glass disk, to which is attached a


delivery tube, without stopcock, about 4 mm in inside diameter and about 50 mm in


length. Pack a pledget of fine glass wool in the base of the tube. Place the specified


volume of


Stationary Phase


in a 100- to 250-mL beaker, add the specified amount of


Solid Support,


and mix to produce a homogeneous, fluffy mixture. Transfer this


mixture to the chromatographic tube, and tamp, using gentle pressure, to obtain a


uniform mass. If the specified amount of


Solid Support


is more than 3 g, transfer the


mixture to the column in portions of approximately 2 g, and tamp each portion. If the


assay or test requires a multisegment column, with a different


Stationary Phase


specified for each segment, tamp after the addition of each segment, and add each


succeeding segment directly to the previous one.



If a solution of the analyte is incorporated in the


Stationary Phase,


complete the


quantitative transfer to the chromatographic tube by scrubbing the beaker used for the


preparation of the test mixture with a mixture of about 1 g of


Solid Support


and several


drops of the solvent used to prepare the test solution.


Pack a pledget of fine glass wool above the completed column packing. The


Mobile


Phase


flows through a properly packed column as a moderate stream or, if


reverse- phase chromatography is applied, as a slow trickle.


Procedure



Transfer the


Mobile Phase


to the column space above the column


packing, and allow it to flow through the column under the influence of gravity. Rinse


the tip of the chromatographic column with about 1 mL of


Mobile Phase


before each


change in composition of


Mobile Phase


and after completion of the elution. If the


analyte is introduced into the column as a solution in the


Mobile Phase,


allow it to


pass completely into the column packing, then add


Mobile Phase


in several small


portions, allowing each to drain completely, before adding the bulk of the


Mobile


Phase.


Where the assay or test requires the use of multiple chromatographic columns


mounted in series and the addition of


Mobile Phase


in divided portions is specified,


allow each portion to drain completely through each column, and rinse the tip of each


with


Mobile Phase


prior to the addition of each succeeding portion.




GAS CHROMATOGRAPHY




The distinguishing features of gas chromatography are a gaseous mobile phase and a


solid or immobilized liquid stationary phase. Liquid stationary phases are available in


packed or capillary columns. In the packed columns, the liquid phase is deposited on


a finely divided, inert solid support, such as diatomaceous earth, porous polymer, or


graphitized carbon, which is packed into a column that is typically 2 to 4 mm in internal


diameter and 1 to 3 m in length. In capillary columns, which contain no packing, the


liquid phase is deposited on the inner surface of the column and may be chemically


bonded to it. In gas-solid chromatography, the solid phase is an active adsorbent,


such as alumina, silica, or carbon, packed into a column. Polyaromatic porous resins,


which are sometimes used in packed columns, are not coated with a liquid phase.


When a vaporized compound is introduced into the carrier gas and carried into the


column, it is partitioned between the gas and stationary phases by a dynamic


countercurrent distribution process. The compound is carried down the column by the


carrier gas, retarded to a greater or lesser extent by sorption and desorption on the


stationary phase. The elution of the compound is characterized by the partition ratio,


k


?


,


a dimensionless quantity also called the capacity factor (see


Glossary of Symbols


for the definition of symbols). It is equivalent to the ratio of the time required for the


compound to flow through the column (the retention time) to the elution time of an


unretained compound. The value of the capacity factor depends on the chemical


nature of the compound, the nature, amount, and surface area of the liquid phase, the


column temperature, and the gas flow rate. Under a specified set of experimental


conditions, a characteristic capacity factor exists for every compound. Separation by


gas chromatography occurs only if the compounds concerned have different capacity


factors.


Apparatus



A gas chromatograph consists of a carrier gas source, an injection port,


column, detector, and recording device. The injection port, column, and detector are


temperature-controlled. The typical carrier gas is helium, nitrogen, or hydrogen,


depending on the column and detector in use. The gas is supplied from a


high-pressure cylinder or high-purity gas generator and passes through suitable


pressure-reducing valves and a flow meter to the injection port and column.


Compounds to be chromatographed, either in solution or as gases, are injec


ted into


the gas stream at the injection port. Depending upon the configuration of the


apparatus, the test mixture may be injected directly into the column or be vaporized in


the injection port and mixed into the flowing carrier gas prior to entering the column.



Once in the column, compounds in the test mixture are separated by virtue of


differences in their capacity factors, which in turn depend upon vapor pressure and


degree of interaction with the stationary phase. The capacity factor, which governs


resolution, retention times, and column efficiencies of components of the test mixture,


is also temperature-dependent. The use of temperature-programmable column ovens


takes advantage of this dependence to achieve efficient separation of compounds


differing widely in vapor pressure.


As resolved compounds emerge separately from the column, they pass through a


differential detector, which responds to the amount of each compound present. The


type of detector to be used depends upon the nature of the compounds to be


analyzed and is specified in the individual monograph. Detectors are heated to


prevent condensation of the eluting compounds.


Detector output is recorded as a function of time, producing a chromatogram, which


consists of a series of peaks on a time axis. Each peak represents a compound in the


vaporized test mixture, although some peaks may overlap. The elution time is a


characteristic of an individual compound; and the instrument response, measured as


peak area or peak height, is a function of the amount present.


Injectors



Sample injection devices range from simple syringes to fully programmable


automatic injectors. The amount of sample that can be injected into a capillary column


without overloading is small compared to the amount that can be injected into packed


columns, and may be less than the smallest amount that can be manipulated


satisfactorily by syringe. Capillary columns, therefore, often are used with injectors


able to split samples into two fractions, a small one that enters the column and a large


one that goes to waste. Such injectors may be used in a


splitless mode


for analyses of


trace or minor components.


Purge and trap injectors are equipped with a sparging device by which volatile


compounds in solution are carried into a low-temperature trap. When sparging is


complete, trapped compounds are desorbed into the carrier gas by rapid heating of


the temperature-programmable trap.


Headspace injectors are equipped with a thermostatically controlled sample heating


chamber. Solid or liquid samples in tightly closed containers are heated in the


chamber for a fixed period of time, allowing the volatile components in the sample to


reach an equilibrium between the nongaseous phase and the gaseous or headspace


phase.


After this equilibrium has been established, the injector automatically introduces a


fixed amount of the headspace in the sample container into the gas chromatograph.


Columns



Capillary columns, which are usually made of fused silica, are typically 0.2


to 0.53 mm in internal diameter and 5 to 60 m in length. The liquid or stationary phase,


which is sometimes chemically bonded to the inner surface, is 0.1 to 1.0 ?


m thick,


although nonpolar stationary phases may be up to 5 ?


m thick. A list of liquid phases in


current use is given in the section


Chromatographic Reagents.




Packed columns, made of glass or metal, are 1 to 3 m in length with internal


diameters of 2 to 4 mm. Those used for analysis typically are porous polymers or solid


supports with liquid phase loadings of about 5% (w/w). High-capacity columns, with


liquid phase loadings of about 20% (w/w), are used for large test specimens and for


the determination of low molecular weight compounds such as water. The capacity


required influences the choice of solid support.


Supports for analysis of polar compounds on low-capacity, low-polarity liquid phase


columns must be inert to avoid peak tailing. The reactivity of support materials can be


reduced by silanizing prior to coating with liquid phase. Acid-washed, flux-calcined


diatomaceous earth is often used for drug analysis. Support materials are available in


various mesh sizes, with 80- to 100-mesh and 100- to 120-mesh being most


commonly used with 2- to 4-mm columns. Supports and liquid phases are listed in the


section


Chromatographic Reagents.




Retention time and the peak efficiency depend on the carrier gas flow rate; retention


time is also directly proportional to column length, while resolution is proportional to


the square root of the column length. For packed columns, the carrier gas flow rate is


usually expressed in mL per minute at atmospheric pressure and room temperature. It


is measured at the detector outlet with a flowmeter while the column is at operating


temperature. The linear flow rate through a packed column is inversely proportional to


the square of the column diameter for a given flow volume. Flow rates of 60 mL per


minute in a 4-mm column and 15 mL per minute in a 2-mm column give identical


linear flow rates and thus similar retention times. Unless otherwise specified in the


individual monograph, flow rates for packed columns are about 30 to 60 mL per


minute. For capillary columns, linear flow velocity is often used instead of flow rate.


This is conveniently determined from the length of the column and the retention time


of a dilute methane sample, provided a flame-ionization detector is in use. At high


operating temperatures there is sufficient vapor pressure to result in a gradual loss of


liquid phase, a process called bleeding.

-


-


-


-


-


-


-


-



本文更新与2021-02-11 17:27,由作者提供,不代表本网站立场,转载请注明出处:https://www.bjmy2z.cn/gaokao/637811.html

(621)CHROMATOGRAPHY《色谱法》USP31的相关文章