-
ASTMD149
美标标准固体绝缘材料电压击穿的实验方法
Standard Test Method for
Dielectric Breakdown Voltage and Dielectric
Strength of Solid Electrical Insulating
Materials at Commercial Power
Frequencies
固体电绝缘材料在工业用电频率下的
电压击穿和介电强度的实验
方法
1.
Scope*
1.1 This test method
covers procedures for the determination
of dielectric
strength of
solid insulating materials at
commercial power frequencies, under
specified conditions.2,3
1.2 Unless otherwise specified, the
tests shall be made at 60Hz. However, this
test method is suitable for use at
anyfrequency from 25 to 800 Hz. At
frequencies above 800 Hz,dielectric
heating is a potential problem.
1.3 This test method is intended to be
used in conjunction with any ASTM
standard or other document that refers
to this test method. References to this
document need to specify the particular
options to be used (see 5.5).
1.4 It is suitable for use at various
temperatures, and in any suitable gaseous
or liquid surrounding
medium.
1.5 This test method
is not intended for measuring the dielectric
strength of
materials that are fluid
under the conditions of test.
1.6 This test method is not intended
for use in determining intrinsic dielectric
strength,direct-voltage dielectric
strength,or thermal failure under electrical
stress (see Test
MethodD3151).
1.7 This test
method is most commonly used to determine the
dielectric
breakdown voltage through
the thickness of a test specimen (puncture). It is
also suitable for use to determine
dielectric breakdown voltage along the
interface between a solid specimen and
a gaseous or liquid surrounding
medium
(flashover). With the addition of instructions
modifying Section 12, this
test method
is also suitable for use for proof
testing.
1.8 This test
method is similar to IEC Publication 243-1. All
procedures in this
method are included
in IEC 243-1. Differences between this method and
IEC
243-1 are largely
editorial.
1.9
This standard does not purport to
address all of the safety concerns, if any,
associated with its use. It is the
responsibility of the user of this standard to
establish appropriate safety and health
practices and determine the
applicability of regulatory limitations
prior to use.
Specific hazardstatements
are given in Section 7. Also see
6.4.1.
ASTMD149
美标
标准固体绝缘材料电压击穿的实验方法
(
三
)
3.1 Definitions:
3.1.1dielectric breakdown voltage
(electric breakdown voltage),
n
—
the
potential
difference
at
which
dielectric
failureoccurs
under
prescribed
conditions in an electrical
insulatingmaterial located between two
electrodes. (See also
Appendix
X1
.)
3.1.1.1
Discussion
—
The term
dielectric breakdown voltage
is sometimes
shortened to
“breakdown voltage.”
3.1.2dielectric failure (under test),
n
—
an event that is evidenced
by
an increase in conductance in the
dielectric under test limiting the
electric field that can be
sustained.
3.1.3dielectric strength,
n
—
the voltage gradient at
which dielectric
failure of the
insulating material occurs under specific
conditions of
test.
3.1.4 electric strength,
n
—
see dielectric
strength.
3.1.4.1
Discussion
—Internationally, “electric
strength” is used
almost
universally.
3.1.5flashover,
n
—
a disruptive electrical
discharge at the surface of
electrical
insulation
or
in
the
surrounding
medium,
which
may
or
may
not
cause
permanent damage to the insulation.
3.1.6 For definitions of other terms
relating to solid insulating
materials,
refer to Terminology
D1711
.
ASTMD149
美标标准固体绝缘材料电压击穿的实验方法
(
四
)
4. Summary of Test Method
4.2 Most commonly, the test voltage is
applied using simple test electrodes on
opposite faces of specimens. The
options for the specimens are that they be
molded or cast, or cut from flat sheet
or plate. Other electrode and specimen
configurations are also suitable for
use to accommodate the geometry of the
sample material, or to simulate a
specific application for which the material is
being evaluated.
4.1 Alternating voltage at a commercial
power frequency (60 Hz, unless
otherwise specified) is applied to a
test specimen. The voltage is
increased
from zero or
from a level
well below the
breakdown
voltage, in
one
of
three
prescribed
methods
of
voltage
application,
until
dielectric
failure of the test specimen
occurs.
5. Significance and
Use
5.1 The dielectric
strength of an electrical insulating material is a
property of interest
for
any application
where an
electrical field will
be present. In
many cases the dielectric strength of a material
will be
the determining factor in the
design of the apparatus in which it is to
be used.
5.2
Tests made as specified herein are suitable for
use to provide part of the
information
needed for determining suitability of a material
for a given
application; and also, for
detecting changes or deviations from normal
characteristics
resulting from processing variables,
aging conditions, or other
manufacturing or environmental
situations. This test method is useful for
process control, acceptance or research
testing.
5.3 Results
obtained by this test method can seldom be used
directly to determine
the dielectric
behavior of a material in an actual application.
In most cases it is
necessary that
these results be evaluated by comparison with
results obtained
from other functional
tests or from tests on other materials, or both,
in order to
estimate their significance
for a particular material.
5.4 Three methods for voltage
application are specified in Section
12
:
Method A,
Short-Time Test; Method B, Step-by- Step Test; and
Method C,
Slow Rate-of-Rise Test.
Method A is the most commonly-used test for
quality-control tests. However,the
longer-time tests, Methods B and C,
which usually will give lower test
results, will potentially give more
meaningful
results
when
different
materials
are
being
compared
with
each
other.
If
a
test
set
with
motor-
driven
voltage
control
is
available,
the
slow
rate-of-rise
test
is
simpler
and
preferable
to
the
step-by-step
test.
The results obtained
from Methods B and C are comparable to each
other.
5.5 Documents
specifying the use of this test method shall also
specify:
5.5.1 Method of
voltage application,
5.5.2
Voltage rate-of-rise, if slow rate-of-rise method
is specified,
5.5.3 Specimen
selection, preparation, and
conditioning,
5.5.4
Surrounding medium and temperature during
test,
5.5.5
Electrodes,
5.5.6 Wherever
possible, the failure criterion of the
currentsensing
element, and
5.5.7 Any desired deviations from the
recommended procedures as given.
5.6 If any of the requirements listed
in
5.5
are missing from the
specifying
document,
then
the
recommendations
for
the
several
variables
shall be
followed.
5.7 Unless the
items listed in
5.5
are
specified, tests made with such
inadequate
reference
to
this
test
method
are
not
in
conformance
with
this
test
method.
If
the
items
listed
in
5.5
are
not
closely
controlled
during
the
test,
it
is
possible
that
the
precisions
stated
in
15.2
and
15.3
will
not be
obtained.
5.8
Variations
in
the
failure
criteria
(current
setting
and
response
time)
of the current sensing
element significantly affect the test
results.
5.9
Appendix
X1
.
contains
a
more
complete
discussion
of
the
significance
of dielectric
strength tests.
ASTMD149
美标标准固体绝缘材料电压击穿的实验方法
(
六
)
6.
Apparatus
6.1.3
The controls on the variable low-voltage source
shall be capable of
varying the supply
voltage and the resultant test voltage smoothly,
uniformly,
and without overshoots or
transients, in accordance with 12.2. Do not allow
the
peak voltage to exceed 1.48 times
the indicated rms test voltage under any
circumstance. Motor-driven controls are
preferable for making short-time (see
12.2.1) or slow-rate-of-rise (see
12.2.3) tests.
6.1.4 Equip
the voltage source with a circuit-
breaking
device that will
operate
within three cycles. The device
shall
disconnect the
voltage-source equipment
from the
power
service and protect it
from overload as a result of specimen
breakdown causing an overload of the
testing apparatus. If
prolonged current
follows
breakdown it will result in unnecessary
burning of the test specimens,
pitting of the electrodes,
and contamination of any liquid
surrounding medium.
6.1.5 It
is important for the circuit-breaking device to
have an
adjustable
current-
sensing
element
in
the
step-up
transformer
secondary,
to
allow
for
adjustment
consistent
with
the
specimen
characteristics
and
arranged to sense specimen current. Set
the sensing element to respond
to
a
current
that
is
indicative
of
specimen
breakdown
as
defined
in
12.3.
6.1.6 The
current setting is likely to have a significant
effect on the
test results. Make the
setting high enough that transients, such as
partial discharges, will not trip the
breaker but not so high that
excessive
burning
of
the
specimen,
with
resultant
electrode
damage,
will
occur on breakdown. The optimum current
setting is not the same for all
specimens and depending upon the
intended use of the material and the
purpose
of
the
test,
it
is
often
desirable
to
make
tests
on
a
given
sample
at more than one
current setting. The electrode area is likely to
have
a significant effect upon the
choice of current setting.
6.1.7 It is possible that the specimen
current-sensing element will be
in
the
primary
of
the
step-up
transformer.
Calibrate
the
current-sensing
dial in terms of specimen
current.
6.1.8 Exercise care
in setting the response of the current
control. If
the control is
set
too high,
the
circuit will not
respond when breakdown
occurs; if set too low, it is
possible
that it will
respond to leakage
currents,
capacitive
currents,
or
partial
discharge
(corona)
currents
or,
when
the sensing element
is
located in the primary, to the step-up
transformer
magnetizing current.
6.2
Voltage
Measurement
—
A voltmeter must
be provided for measuring the
rms
test
voltage.
If
a
peak-reading
voltmeter
is
used,
divide
the
reading
by
=2
to
get
rms
values.
The
overall
error
of
the
voltage-measuring
circuit
shall
not
exceed
5
%
of
the
measured
value.
In
addition,
the
response
time
of
the
voltmeter
shall
be
such
that
its
time
lag
will
not
be
greater
than
1 % of full scale at any rate-of-rise
used.
6.2.1 Measure the
voltage using a voltmeter or potential
transformer
connected
to
the
specimen
electrodes,
or
to
a
separate
voltmeter
winding,
on the test transformer, that
is
unaffected by the step-up
transformer
loading.
6.2.2
It
is
desirable
for
the
reading
of
the
maximum
applied
test
voltage
to
be retained on the voltmeter after breakdown
so
that the breakdown
voltage can be accurately read and
recorded.
ASTMD149
美标标准固体绝缘材料电压击穿的实验方法
(
七
)
点击次数:
138
发布时间:
2010-7-13 11:01:45
6.3
Electrodes
—
For a
given specimen configuration, it is
possible that the
dielectric
breakdown voltage will vary
considerably,
depending upon
the
geometry and placement of
the
test electrodes. For
this reason it is important
that the
electrodes
to be used be
described when specifying this test
method, and that they be described in
the report.
6.3.1 One of the
electrodes listed in
Table 1
shall be
specified by the document
referring to this test method. If
no
electrodes have been
specified, select an
applicable one
from
Table 1
, or
use other electrodes mutually acceptable to
the
parties
concerned when the standard electrodes cannot be
used
due to the
nature or configuration of the material
being tested.
See references
in
Appendix X2
for examples
of some special
electrodes.
In any event the
electrodes must be
described in the
report.
6.3.2 The
electrodes of Types 1 through 4 and Type 6
of
Table 1
shall
be in
contact with the test specimen
over the
entire flat area of
the electrodes.
6.3.3 The
specimens tested using Type 7 electrodes shall
be
of such size
that all portions of the specimen will be within
and no less than 15
mm from the edges
of the electrodes during test. In most cases,
tests using
Type 7 electrodes are made
with the plane of the electrode surfaces in a
-
-
-
-
-
-
-
-
-
上一篇:复试英语机械专业英语
下一篇:高中英语外研版选修八单词表84601