-
Stopped-Flow Technique
Perhaps
the most frequently used rapid kinetics techniques
is stopped-flow. Small volumes of
solutions are rapidly driven from
syringes into a high efficiency mixer to initiate
a fast
reaction. The resultant reaction
volume then displaces the contents of an
observation cell thus
filling it with
freshly mixed reagents. The volume injected is
limited by the stop syringe
which
provides the “stopped
-
flow”.
Just prior to stopping, a steady state flow is
achieved.
The solution entering the
flow cell is only milliseconds old. The age of
this reaction volume
is also known as
the dead time of the stopped-flow system. As the
solution fills the stopping
syringe,
the plunger hits a block, causing the flow to be
stopped instantaneously. Using
appropriate techniques, the kinetics of
the reaction can be measured in the cell.
Traditionally
stopped-flow has lent itself to UV/Vis detection
methods
–
both by absorption
and fluorescence spectroscopy and these
are still the most common methods employed. TgK
Scientific offer a comprehensive range
of stopped-flow equipment from simple stopped-flow
accessories that can be used with
existing absorbance and fluorescence spectrometers
to
complete, self-contained computer
controlled stopped-flow systems with dedicated
acquisition and analysis software.
Mixing options for stopped-
flow include multi-mixing (also known as double
mixing or
sequential mixing) where more
than two solutions are mixed together, variable
ratio mixing
using different size
syringes and microvolume mixing for applications
where only very small
quantities of
reactants are available.
The environment of the reagents is also
extremely important and so equipment is available
for anaerobic applications, sub-zero
temperature and high pressure stopped-flow
applications.
An extensive range of
software packages is also available for the
interpretation and analysis
of the
kinetic data. In addition to equipment for
absorbance and fluorescence measurements,
Hi-Tech offers dedicated stopped-flow
systems for circular dichroism, conductivity and
FT-
IR measurement.
Stopped-
Flow - A technique for studying rapid chemical
kinetics
Stopped-flow is one of a
number of techniques used to study the kinetics of
reactions in
solution. In the simplest
form of the technique, two reactant solutions are
rapidly mixed by
being forced into a
mixing chamber, and then through an observation
cell. At some point in
time, the flow
is suddenly stopped, and the reaction monitored
using a suitable spectroscopic
probe,
such as absorbance, fluorescence or circular
dichroism. The change in the
spectroscopic signal as a function of
time is recorded.
Figure 1 shows the
schematic of a typical stopped-flow apparatus. The
stopping mechanism
in this example is a
stop-syringe. The flow fills the stop-syringe,
until the plunger hits the
trigger-
switch. This simultaneously stops the flow and
starts the data acquisition.
The rate
constants which define the reaction kinetics, can
be measured by fitting the data
using a
suitable model.
The performance of a
stopped-flow instrument is determined to a large
extent by the dead-
time. This is
defined as the minimum time after the reactant
have mixed that the observation
starts.
The dead-time is essentially the age of the
reaction as it enters the observation cell. The
limiting factor in the dead-time of a
particular stopped-flow apparatus is determined by
the
distance between the mixer and the
cell, and the final velocity of the flow at the
instant the
flow is stopped. Another
factor which can affect the dead-time is the
efficiency of the mixer.
Typically,
dedicated stopped-flow instruments can achieve
dead-times in the region of a
millisecond. Using ultra-small
observation cells, dead times of less than 0.5ms
can be
achieved. Typically however, the
true experimental limitation is the rate of mixing
of
solutions of the two reactants.
-
-
-
-
-
-
-
-
-
上一篇:大学新视野英语第四册第一单元网络试题
下一篇:新托福TPO24阅读原文及译文(三)