关键词不能为空

当前您在: 主页 > 英语 >

Stopped flow technique

作者:高考题库网
来源:https://www.bjmy2z.cn/gaokao
2021-02-10 03:20
tags:

-

2021年2月10日发(作者:diss什么意思)


Stopped-Flow Technique


Perhaps the most frequently used rapid kinetics techniques is stopped-flow. Small volumes of


solutions are rapidly driven from syringes into a high efficiency mixer to initiate a fast


reaction. The resultant reaction volume then displaces the contents of an observation cell thus


filling it with freshly mixed reagents. The volume injected is limited by the stop syringe


which provides the “stopped


-


flow”. Just prior to stopping, a steady state flow is achieved.


The solution entering the flow cell is only milliseconds old. The age of this reaction volume


is also known as the dead time of the stopped-flow system. As the solution fills the stopping


syringe, the plunger hits a block, causing the flow to be stopped instantaneously. Using


appropriate techniques, the kinetics of the reaction can be measured in the cell.




Traditionally stopped-flow has lent itself to UV/Vis detection methods



both by absorption


and fluorescence spectroscopy and these are still the most common methods employed. TgK


Scientific offer a comprehensive range of stopped-flow equipment from simple stopped-flow


accessories that can be used with existing absorbance and fluorescence spectrometers to


complete, self-contained computer controlled stopped-flow systems with dedicated


acquisition and analysis software.



Mixing options for stopped- flow include multi-mixing (also known as double mixing or


sequential mixing) where more than two solutions are mixed together, variable ratio mixing


using different size syringes and microvolume mixing for applications where only very small


quantities of reactants are available.




The environment of the reagents is also extremely important and so equipment is available


for anaerobic applications, sub-zero temperature and high pressure stopped-flow applications.


An extensive range of software packages is also available for the interpretation and analysis


of the kinetic data. In addition to equipment for absorbance and fluorescence measurements,


Hi-Tech offers dedicated stopped-flow systems for circular dichroism, conductivity and FT-


IR measurement.


Stopped- Flow - A technique for studying rapid chemical kinetics


Stopped-flow is one of a number of techniques used to study the kinetics of reactions in


solution. In the simplest form of the technique, two reactant solutions are rapidly mixed by


being forced into a mixing chamber, and then through an observation cell. At some point in


time, the flow is suddenly stopped, and the reaction monitored using a suitable spectroscopic


probe, such as absorbance, fluorescence or circular dichroism. The change in the


spectroscopic signal as a function of time is recorded.


Figure 1 shows the schematic of a typical stopped-flow apparatus. The stopping mechanism


in this example is a stop-syringe. The flow fills the stop-syringe, until the plunger hits the


trigger- switch. This simultaneously stops the flow and starts the data acquisition.


The rate constants which define the reaction kinetics, can be measured by fitting the data


using a suitable model.


The performance of a stopped-flow instrument is determined to a large extent by the dead-


time. This is defined as the minimum time after the reactant have mixed that the observation


starts. The dead-time is essentially the age of the reaction as it enters the observation cell. The


limiting factor in the dead-time of a particular stopped-flow apparatus is determined by the


distance between the mixer and the cell, and the final velocity of the flow at the instant the


flow is stopped. Another factor which can affect the dead-time is the efficiency of the mixer.


Typically, dedicated stopped-flow instruments can achieve dead-times in the region of a


millisecond. Using ultra-small observation cells, dead times of less than 0.5ms can be


achieved. Typically however, the true experimental limitation is the rate of mixing of


solutions of the two reactants.

-


-


-


-


-


-


-


-



本文更新与2021-02-10 03:20,由作者提供,不代表本网站立场,转载请注明出处:https://www.bjmy2z.cn/gaokao/625991.html

Stopped flow technique的相关文章

  • 余华爱情经典语录,余华爱情句子

    余华的经典语录——余华《第七天》40、我不怕死,一点都不怕,只怕再也不能看见你——余华《第七天》4可是我再也没遇到一个像福贵这样令我难忘的人了,对自己的经历如此清楚,

    语文
  • 心情低落的图片压抑,心情低落的图片发朋友圈

    心情压抑的图片(心太累没人理解的说说带图片)1、有时候很想找个人倾诉一下,却又不知从何说起,最终是什么也不说,只想快点睡过去,告诉自己,明天就好了。有时候,突然会觉得

    语文
  • 经典古训100句图片大全,古训名言警句

    古代经典励志名言100句译:好的药物味苦但对治病有利;忠言劝诫的话听起来不顺耳却对人的行为有利。3良言一句三冬暖,恶语伤人六月寒。喷泉的高度不会超过它的源头;一个人的事

    语文
  • 关于青春奋斗的名人名言鲁迅,关于青年奋斗的名言鲁迅

    鲁迅名言名句大全励志1、世上本没有路,走的人多了自然便成了路。下面是我整理的鲁迅先生的名言名句大全,希望对你有所帮助!当生存时,还是将遭践踏,将遭删刈,直至于死亡而

    语文
  • 三国群英单机版手游礼包码,三国群英手机单机版攻略

    三国群英传7五神兽洞有什么用那是多一个武将技能。青龙飞升召唤出东方的守护兽,神兽之一的青龙。玄武怒流召唤出北方的守护兽,神兽之一的玄武。白虎傲啸召唤出西方的守护兽,

    语文
  • 不收费的情感挽回专家电话,情感挽回免费咨询

    免费的情感挽回机构(揭秘情感挽回机构骗局)1、牛牛(化名)向上海市公安局金山分局报案,称自己为了挽回与女友的感情,被一家名为“实花教育咨询”的情感咨询机构诈骗4万余元。

    语文