Summary



Description/


描述



The product temperature sensor in Glattfluid bed drier cannot be calibrated before removing from the


system, discrepancy was initiated according to SOP-0102619 section 6.1.7


initiated when a calibration fails or no as found readings can be taken before the sensor removed


from the calibration schedule


Glatt


制粒机里的产品温度传感器在从系 统删除前无法做校验,根据


SOP-0102619


步骤


6.1.7


“校准失败


或当仪表从校准计划中 删除无法获得前校数据时,必须启动差异处理程序”启动偏差。




1. Deviation/ Actions /


偏差行动



Production staffobserved the product temperature sensor wasbroken on 2015.11.24, maintenance


team replaced the broken product temperature sensor with a new


improved


one on 2015.11.26. Work


order refers to attachment TW1118464-01.


生产部员工


2015.11.2 4


日产品温度传感器损坏,维修团队于


2015.11.26< /p>


日更换了新的经过改进的产品温


度传感器。工单见附件

< p>
TW1118464-01





2. Causal Analysis /


因果分析




5 Why’s tool w


as


used to do the cause analysis.5 Why’s diagram refers to attachment


TW1118464-02.


用了


5 Why



s


工具进行原因分析。


5 Why’s


示意图参见附件


TW1118 464-02





1st



Why the temperature sensor could not be calibrated? /


为什么温度探头无法进行校验?




At about 23:32 on 2015.11.24 after completion the batch of Xeldoa SH2105



end at 22:50 on


2015.11.24



, alarm 101-7



Product Temperature FBD(T202010)- Broken Wire



occurred, operator


checked the alarm and found that the part of FBD product temperature sensor holderinstalled out of


the chamber was broken. Electrical engineer was informed and confirmed that the sensor was broken.


2015.11.24


晚上


23



32


左右在结束希罗达


SH2105


批次后


(2015.11.24, 22:50


结束),


Glatt


系统发生


报警< /p>


101-7


“物料温度(


T202010



-


断线”,员工检查发现安装在< /p>


FBD


产品温度探头的腔体外部分连


杆处 断裂。通知了电工,电工确认该温度探头损坏。




Calibration team checked the sensor and evaluated the sensor could notbe calibratedas the


thermocouple melting point of the sensor felloff and no output signal detected from sensor.


校验小组对探头进行了检查,评估该探头无法进行校准,因为探头热电 偶熔点掉落且没有检测到信号输


出。




The broken part was installed outside of the FBD chamber, picture of broken product temperature


sensor refers to attachment broken point was marked with arrow


损害的部 分在


FBD


墙体外面,损坏的产品温度探头图见附件

< p>
TW1118464-03


。箭头标注了断裂点。




Above all,damage of FBD product temperature sensor caused that the sensor cannot be calibrated.


综上,


FBD


的产品温度探头损坏导致该探头无法进行校准。




2nd- Why the FBD product temperature sensor was broken? /


为什么


FBD


温度探头被损坏?




By checking the broken product temperature sensor, the thickness of temperature sensor


holderbroken point was about 1mm,made of 316L stainless. The end of the holderwas a junction box


used to install temperature transmitter, and it is a bit heavy.


通过查看损坏的产品温度探头,


Glatt


产品温度传感器连杆的断裂处壁厚约为


1


毫 米,为


316L


不锈钢材


质,其后部 为接线盒,此接线盒用于安装温度变送器,较重。




It is possible the thinner stainless of temperature sensor holder combining the heavier junction box in


the end led to mechanical fatigue over time and caused the holder broke in the holder broke,


the junction box fell down under the gravity, pulled the temperature sensor and caused that the


thermocouple melting point of the sensor


fell off.


可能由于温度传感器连杆 的不锈钢厚度较薄,并且其后端的接线盒较重,因此使用中容易造成材质的机


械疲劳,发 生断裂。连杆断裂后,接线盒由于重力作用下坠,拉扯温度探头,导致温度探头热电偶熔点


掉落。




The broken temperature sensor passed the calibration before installed on the FBD, the calibration


certificate refers to TW1118464-05(TT-00142-PRO). After the discrepancy occurred, calibration team


initiated the calibration decommission for it. See attachment TW1118464-05 ( TT-00142-PROCCAF-


decommission).


已损坏的温度探头在安装到制粒机前校准合格,校准证书见附件


TW1118 464-05 (TT-00142-PRO)


。此


偏差发生后 ,校验小组对该探头进行了报废,详见附件


TW1118464-05 (TT-001 42-PROCCAF-


decommission)





Similar case occurred in July 2015, the deviation report number was TW1062179, the root cause was


same as TW1118464. Following remediation action was defined in TW1062179 followed as post


closure action TW1069248: contact Supplier, strength the temperature sensor holder, add thickness


up to 2mm, and replace the wiring in the junction box in the end by 2 meter of flexible cord


supplier responded it would take time to change the design of the sensor , the target


completion date for this action is 2015.12.31.


20 15.07


月发生过类似事件,详见偏差


TW1062179< /p>


。偏差


TW1062179


制定了后续行 动以关闭后行动


TW1069248


跟踪:联系供应商,加厚温 度传感器连杆壁厚至


2


毫米,同时后端接线盒接线方式改为直< /p>



2


米软线引出。供应商回复因为改这个 感应器的设计需要花一定的时间,所以该行动的预计完成日期



2015.12.31





When the discrepancy (TW1118464) occurred on 2015.11.24, maintenance team just received the


new sensor and the calibration was not completed yet.


本偏差(


TW1118464


)在


2015.1 1.24


日发生时,维护小组刚收到新的感应器,还没有做校验。




Cause conclusion/


原因总结



The thickness of temperature sensor holderwas about 1mm, made of 316L stainless. The end of the


holderwas a junction box used to install temperature transmitter, and it is a bit is possible the


thinner stainless of temperature sensor holder combining the heavier junction box in the end led to


mechanical fatigue over time and caused the holder broke in the holder broke, the junction


box fell down due to gravity, pulled the temperature sensor and caused that the


thermocouple


melting point of the sensor


fell off, and the calibration cannot be conducted.


Glatt< /p>


产品温度传感器连杆的断裂处壁厚约为


1


毫米,


316L


不锈钢材质,其后部为接线盒,此接线盒< /p>


用于安装温度变送器,较重。推测由于温度传感器连杆的不锈钢厚度较薄,并且其后端的接 线盒较重,


因此使用中随着时间的推移容易造成材质的机械疲劳,发生断裂。连杆断裂后 ,接线盒由于重力作用下


坠,拉扯温度探头,导致温度探头热电偶熔点掉落,无法进行校 准。




Recurring discrepancy assessment/


重复发生的差异的评估




Timeframe: From 2013.11.19(gDMS in Trackwise go-life in Shanghai site) till 2015.12.09



Date


Closed



时间框架:查询的时间段从


2013.11.19(Trackwise

< p>


gDMS


在上海上线的日期

)



2015.12.09


(关闭


时间)



Assigned site/


工厂


: Shanghai


Responsible Operational Area/


操作区域:


production


Equipment ID /


设备编号


: SH_SHiP-Granulation system


Searching per field on Rec text with



all these words

< br>)


/


在每个可输入文字的区域搜索关键词


(


所有这些



):Produ ct Temperature Sensor



Result: two records were found, TW864400 & TW1062179. TW864400 was related to LOD of


Cellcept Cap out of acceptance 1062179 was same as the newdiscrepancy. So the trend


status is recurring.


结 果:搜索到


2


个记录,


TW86440 0 & TW1062179



TW864400


是关于骁悉胶囊干燥失重超限。其中


TW1062179

和此差异一致,因此趋势状态为“重复”。



-


-


-


-


-


-


-


-



本文更新与2021-02-09 00:29,由作者提供,不代表本网站立场,转载请注明出处:https://www.bjmy2z.cn/gaokao/617885.html

偏差处理案例的相关文章

  • 余华爱情经典语录,余华爱情句子

    余华的经典语录——余华《第七天》40、我不怕死,一点都不怕,只怕再也不能看见你——余华《第七天》4可是我再也没遇到一个像福贵这样令我难忘的人了,对自己的经历如此清楚,

    语文
  • 心情低落的图片压抑,心情低落的图片发朋友圈

    心情压抑的图片(心太累没人理解的说说带图片)1、有时候很想找个人倾诉一下,却又不知从何说起,最终是什么也不说,只想快点睡过去,告诉自己,明天就好了。有时候,突然会觉得

    语文
  • 经典古训100句图片大全,古训名言警句

    古代经典励志名言100句译:好的药物味苦但对治病有利;忠言劝诫的话听起来不顺耳却对人的行为有利。3良言一句三冬暖,恶语伤人六月寒。喷泉的高度不会超过它的源头;一个人的事

    语文
  • 关于青春奋斗的名人名言鲁迅,关于青年奋斗的名言鲁迅

    鲁迅名言名句大全励志1、世上本没有路,走的人多了自然便成了路。下面是我整理的鲁迅先生的名言名句大全,希望对你有所帮助!当生存时,还是将遭践踏,将遭删刈,直至于死亡而

    语文
  • 三国群英单机版手游礼包码,三国群英手机单机版攻略

    三国群英传7五神兽洞有什么用那是多一个武将技能。青龙飞升召唤出东方的守护兽,神兽之一的青龙。玄武怒流召唤出北方的守护兽,神兽之一的玄武。白虎傲啸召唤出西方的守护兽,

    语文
  • 不收费的情感挽回专家电话,情感挽回免费咨询

    免费的情感挽回机构(揭秘情感挽回机构骗局)1、牛牛(化名)向上海市公安局金山分局报案,称自己为了挽回与女友的感情,被一家名为“实花教育咨询”的情感咨询机构诈骗4万余元。

    语文